User's Manual Part 3

To set recording parameters:
1) Connect the M.O.L.E. Profiler to the computer. Refer to
Basics>Setup>Communications Setup for more information.
If an instrument is not currently connected to the computer, the default
Demonstration MEGAM.O.L.E.® profiler will be displayed.
2) On the M.O.L.E. menu, click Set Recording Parameters.
The Start and Stop Parameters are optional settings and do not require
configuration.
3) In the Instrument Name text box, type a company, operator, or M.O.L.E. Profiler
name.
4) Enter the Recording interval at which the M.O.L.E. profiler records data points.
The maximum interval for a SuperM.O.L.E.® Gold Profiler is 24 hours (one
data point per day) and the minimum interval is 0.3 second (one data point
every three-tenths of a second). When using a SuperM.O.L.E.® Gold Profiler
with the RF Option the minimum interval is 0.5 (one data point every
five-tenths of a second).