User's Manual Part 3
• Mains Transient Overvoltage Installation Category II (Per IEC 664)
•
Pollution Degree 2 (Normally only dry pollution, but with temporary conductivity caused by condensation)
NOTE:Tests have shown that because of the sensitive nature of the measurement and logic circuits, the
following precautions must be observed:
• Minimize exposure to ESD Events. If the M.O.L.E. Profiler or one of the thermocouples receives an 8kV
electrostatic discharge during the data collection sequence, the M.O.L.E. Profiler may switch itself “OFF”.
To retrieve the data, simply upload the data that was recorded prior to the event. If a new data collection
sequence is started without uploading first, the existing data stored in the M.O.L.E. Profiler will be lost.
• Keep strong electromagnetic fields away. The thermocouple wires serve as an antenna for
electromagnetic radiation. If field strength of 3 volts per meter is present (usually due to close proximity
of
radio transmitters) while the M.O.L.E. Profiler is collecting data, the accuracy of the data may be
compromised. This interference is compounded by using long thermocouple wires, as well as the effect of
the thermocouple wires acting as a “tuned” antenna.
SuperM.O.L.E.® Gold 2 Profiler
INPUTS:
Up to 6 type K ECD Micro-Thermocouples
PHYSICAL DIMENSIONS:
9.41mm x 89mm x 152.4mm (0.37in. x 3.5in. x 6in.)
TEMPERATURE MEASUREMENT RANGE:
-200°C to +1271°C (-328°F to +2322°F)
INTERNAL OPERATING TEMPERATURE RANGE:
-40°C to 85°C (-40°F to 185°F)
Absolute Maximum Warranteed Internal
Temperature:
87°C (188.6°F)* SuperM.O.L.E.® Gold 2 Profiler
automatically stops operating when the internal
temperature exceeds 85°C (185°F)
*WARNING: Exceeding this temperature may permanently damage the equipment!
ACCURACY:
±1°C. Channel to channel compliance within 1°C.
PROGRAMMABLE SAMPLING PERIOD:
0.1 seconds to 24 hours
HARDWARE REAL-TIME CLOCK/CALENDAR:
±1 minute/month at 25°C
OPEN THERMOCOUPLE DETECTION.
All 6 channels
NUMBER OF SAMPLES PER CHANNEL
250,000 total for 6 channels
Internal Power Pack:
POWER PACK CHARGE:
Typical 50 ten-minute profiles.
CHARGE TIME:
8 hours (a 15 min. charge allows one 10min. data run)
EXPECTED POWER PACK LIFE:
300-400 charging cycles.
OPERATING RANGE:
4.2V to 3.0V
Internal Power Pack Charger:
INPUT:
Voltage Rating: 100 to 240 VAC
Current: 200mA(RMS) max. @ 115VAC
80mA(RMS) max. @ 230VAC
Frequency: 50/60Hz
OUTPUT:
5V DC, Load 0A MIN. 1A MAX.
Mains Voltage Fluctuations:
+ 10%