TEMPERATURE MEASUREMENT RANGE:
-129°C to +1250°C (200°F to +2282°F)
INTERNAL OPERATING TEMPERATURE RANGE:
0°C to 65°C (32°F to 150°F)
Absolute Maximum Warranteed Internal
Temperature:
82°C (180°F)* SuperM.O.L.E.® Gold automatically
stops monitoring when the internal temperature
exceeds 80°C (176°F)
*WARNING: Exceeding this temperature may permanently damage the equipment!
ACCURACY:
Within ± 0.1%+1°C. at –73, 100, 250 and 1000°C.
Channel to channel compliance within 1°C.
PROGRAMMABLE SAMPLING PERIOD:
0.1 seconds to 24 hours
HARDWARE REAL-TIME CLOCK/CALENDER:
±1 minute/month at 25°C
OPEN THERMOCOUPLE DETECTION.
All six channels
NUMBER OF SAMPLES PER CHANNEL
5460 total for six channels
Power Pack:
POWER PACK CHARGE:
Typical 50 ten-minute profiles.
CHARGE TIME:
14 hours.
EXPECTED POWER PACK LIFE:
300-400 charging cycles.
OPERATING RANGE:
5.1V to 4.9V (the unit may become unreliable below
4.5V).
Power Pack Charger Model Specifications:
NORTH AMERICA:
ECD Part No. E31-0900-25
Input: 120 V, AC, 50/60 Hz, 2.8 VA Output: 9V DC, 120-200mA, 1.08VA
CONTINENTAL EUROPE:
ECD Part No. E31-0900-21
Input: 230 V, AC, 50/60 Hz, 2.8 VA Output: 9V DC, 120-200mA, 1.08VA
Thermocouple Specifications:
Thermocouples: Type K, Micro-Connector, Glass insulated. (Other insulation types available)
THERMOCOUPLE RESPONSE TIME
Wire Size
Still Air
800/100°F
60 Ft./Sec. Air
800/100°F
Still H2O
200/100°F
.001 in.
.005 in.
.015 in.
0.05 sec.
1.0 sec.
10.0 sec.
0.004 sec.
0.08 sec.
0.80 sec.
0.002 sec.
0.04 sec.
0.40 sec.
Environmental Limitation Specifications:
• Altitude up to 2000 meters
• Maximum relative humidity 80% RH from 0°C to 31°C, decreasing linearly to 17% RH at 50°C.
•
Pollution Degree 2 (Normally only dry pollution, but with temporary conductivity caused by condensation)
Power Pack Charger Only:
• Mains Voltage Fluctuations + 10%