User's Manual Part 3
19. The M.O.L.E.® reports success or failure. If successful, Hyperterminal dispays:
Gain Calibration completed! then hit the Enter key which displays the “?_”. If
failure, repeat ^OC3 as directed in Step 14.
20. Now perform a calibration confirmation. Select Disconnect on the
Hyperterminal Toolbar, disconnect the M.O.L.E.® from the computer and
record several temperature values downloading them into M.O.L.E.® MAP
software to see if they are within the ECD specification. If acceptable, connect
the M.O.L.E.® to the computer again and select Call on the Hyperterminal
Toolbar. Enter: ^OCD then ^OCA.
Manufacturer Specification: +/- 1ºC of reading
Those commands set the calibration date and stores the calibration constants
as archive values:
Your M.O.L.E.® Thermal Profiler is now calibrated.
7.3.4. Constructing a Thermocouple
The following procedures describe how to construct a thermocouple and a shorting plug.
Thermocouple construction:
The following items will be needed to construct a Thermocouple: