User's Manual Part 3
MEGAM.O.L.E.® Help System Guide ♦288♦
5) If desired, configure a Start Parameter such as a threshold temperature or Delay
Points by selecting the associated check box and entering the proper values.
Specifying a threshold temperature “triggers” the recording process when any
active channel reaches the specified temperature and Data Points “trigger” the
M.O.L.E. profiler to start recording when the specified data point is reached in
the process. The actual delay is equal to the Interval times the Pts Dly.
6) If desired, configure a Stop Parameter by selecting the associated check box and
entering the proper value. This parameter is a number of data points the
M.O.L.E. profiler will record during the data run. The maximum number is
dependent on the number of channels turned on.
If a no stop parameter is specified, the M.O.L.E. Profiler will log continuously
and will require the user to manually stop the recording process.
7) Configure sensor information by turning the amount of channels ON or OFF, set
the sensor location description and sensor type.
8) Click the OK command button to send the data to the instrument or Cancel to
quit the command.