User's Manual Part 3

MEGAM.O.L.E.® Help System Guide 356
6.2.2. Calibration Information
Because the MEGAM.O.L.E. profiler and I/O Module are made with precision
components with high temperature stability and tight tolerances; the analog-to-digital
converter remains stable for years. High quality components together with software
algorithms based on the ITS-90* standard for Type K thermocouples have been
provided to yield specified accuracy and long-term stability. Each unit has been tested
at the factory before it is shipped.
ECD recommends the MEGAM.O.L.E. profiler I/O Module is factory re-
calibrated every 6 months when it is being used constantly. If the use is
occasional, a period of no greater than 12 months between calibrations is
recommended.
Depending on use, however, temperature accuracy should be periodically verified using
a suitable temperature standard. Any observed inaccuracies are probably not caused by
calibration error but by any one of a number of other sources, primarily the following:
1) Poor thermocouple connectors or open thermocouples.
2) Using a temperature standard that is inaccurate or is not traceable to the
National Institute of Standards and Technology (NIST, formerly the Nation
Bureau of Standards).
Check the accuracy of your temperature standard and that it is traceable to NIST.
Be sure that you’re using Type K wire connected to the standard. Be sure that
your temperature standard is cold-junction compensated, or use an ice point
reference.
3) Extremely low Power Pack battery charge.
Recharge the Power Pack battery. Refer to topic Basics>Setup>Charging the
Power Pack Battery.
4) Sudden changes or excessive ambient temperatures.
Allow MEGAM.O.L.E. profiler temperature to stabilize for 1/2 hour before
calibration.
If after checking these possible sources of inaccuracy the MEGAM.O.L.E. profiler I/O
Module still needs to be calibrated, contact ECD for the proper calibration procedure
P/N A47-6342-05.
If you need to return the MEGAM.O.L.E. profiler I/O Module for re-calibration, contact
ECD and request an RMA (Return Merchandise Authorization). Refer to section How to
Get Additional Help for contact information.
*ITS-90 (International Temperature Scale of 1990)