Operation Manual
7
STORAGE & MAINTENANCE
• Keep away the unit from direct sunlight, extreme temperatures, humidity or moisture.
• Use a dry, soft cloth to clean the unit, or if desired, use a cloth lightly dampened with water.
• Do not use alcohol, benzene, thinner or other volatile liquids to clean the unit.
• Do not wash or expose the wrist cu to liquid.
• Remove batteries from the unit when not in use for more than 3 months.
SPECIFICATIONS
Measurement method Oscillometric
Measurement range Pressure: 20 – 280 mmHg: Pulse Rate: 40 – 195 beats/min
Accuracy Pressure: +/- 3 mmHg: Pulse Rate: +/- 5%
Ination Pump driven
Pressure detection Semi-conductor
Power supply 4 x 1,5V AA batteries
Auto power o 3 minutes after last operation
Storage environment -20°C to 60°C, 10% to 95% RH
Operation environment 10°C to 40°C, 30% to 85% RH
External dimensions 155 (L) x 130(W) x 110 (H)
Net weight 410g (batteries excluded)
Cu dimensions Arm circumference of 22~34cm
TROUBLESHOOTING
Note: If your problem cannot be solved by the above, consult your store of purchase.
Do not disassemble the unit.
Problem Cause Solution
No display when you press Are the batteries done? Replace with new batteries.
the Power/Start button Are the battery polarities wrong? Re-align batteries with correct polarities.
“Err P” displayed Fails to inate
“Err 1” displayed Deates too rapidly Return for servicing.
“Err 2” displayed Movement during measurement Do not move during measurement
Signal interference
“Err 3” displayed Incorrect results
Check if the cu is properly connected.
Replace the air tube if it is broken.
Remove interfering source e.g. mobile
phones, magnets. Measure again.
The instrument is in line with the EU Medical Products Directive 93/42/EC.
Application part type BF
The appliance should be disposed of according to Regulation 2002/96/EC-WEEE (Waste
Electrical and Electronic Equipment).
Manufactured in accordance with international standards.
Note: Subject to modication without prior notice.
GUARANTEE
EKS guarantees that all defects due to materials or manufacturing faults will be remedied by replacement
or repair for a period of 2 years from the date of purchase. Please keep the purchase receipt and guarantee
card in a safe place. If you have any complaints, please return the product to your dealer with the guarantee
card and the receipt.
0123