Product Manual

CAUTION:
• To reduce the risk of device interference, keep the Eko Core
at least 1 meter away from all RF emitters including Wi routers
and radios.
To reduce the risks associated with infection follow all
cleaning and disinfecting instructions included in this manual.
Establish and follow a cleaning and disinfecting schedule.
To reduce the risks associated with inaccurate data
acquisition store and operate this stethoscope only as
instructed in this manual. Though there is an acoustic (non-
amplied) mode available with this stethoscope, it is highly
recommended that the battery be recharged within thirty
minutes of the LED indicator turning red. Recharge the battery
using only the USB power cord and charger provided with the
device.
DO NOT immerse the stethoscope in a liquid or subject it to
any sterilization processes other than those described in this
manual.
To reduce the risks associated with very strong
electromagnetic elds avoid using the stethoscope near
strong radio frequency (RF) signals or portable and/or mobile
RF devices. If sudden or unexpected sounds are heard, move
away from any radio transmitting antennas. Using accessories,
transducers, and cables not produced by Eko Devices may result
in increased RF emissions or decreased immunity of the Eko
Electronic Stethoscope System.
Please read, understand, and follow all safety information
contained in these instructions prior to using the Eko Electronic
Stethoscope System. It is recommended that these instructions
be retained for future reference.
To reduce the risk associated with an electrical shock do
not use the stethoscope on patients without the analog
stethoscope’s chest piece in place.
Core Model E4 contains a Bluetooth Class 2 wireless data
link. The maximum radio frequency eld strength generated
by the stethoscope is below three volts per meter, a level
that is considered safe to use with other medical devices.
However, audio, video, and other similar equipment may cause
electromagnetic interference. If such devices are encountered