Product Manual
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Eko Electronic Stethoscope System is intended for use in the
electromagnetic environment specied below. The user of the Eko
Electronic Stethoscope System should assure that it is used in such an
environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment-Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
Not
Applicable
Radiated RF
IEC 61000-4-3
3V/m
80 MHz to
2.5 GHz
3 V/m
80 MHz to
2.5 GHz
d = 1.2 √P 80 MHz to
800 MHz
d = 2.3 √P 800 MHz to
2.5 GHz
where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from
xed RF transmitters,
as determined by an
electromagnetic site
survey,
a
should be less
than the compliance level
in each frequency range.
b
Interference may occur in
the vicinity of equipment
marked with the
following symbol:
33
© 2015 Eko Devices, Inc.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment-Guidance
Power
frequency
(50/60 Hz)
magnetic
eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial
magnetic eld or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level