Product Manual

Technical data
HemoPoint
®
H2 User Manual
59
8. Technical data
8.1 HemoPoint
®
H2 analyzer
Measuring
procedure
Optical absorption photometry.
Source
Dual-Color-LED 570 / 880 nm
Dominant
wavelength of
source
1
st
wavelength: 570 ± 5 nm
2
nd
wavelength: 880 ± 10 nm
Spectral half value
width of the source
1
st
wavelength: 15 ± 3 nm
2
nd
wavelength: 50 nm
Receiver
Photodiode 350 – 820 nm
Measuring range
0 – 25.6 g/dL
Sample material
Venous, arterial, or capillary human blood.
Sample carrier
HemoPoint
®
H2 n•x•t
®
Microcuvette.
Sample size
8 µL
Linearity
0- 20.0 g/dL ± 0.3 g/dL
>20.0 g/dL ± 0.7 g/dL
Measuring time
Approximately 30 – 60s.
Memory (data
storage)
Up to 4000 results, including date and
time.
Dimensions
(LxWxH)
6.25 in. x 6.5 in. x 2.5 in.