5
Move the slider to the CREAM / AIR interface. Look through the
transparent slider and position the vertical black line on the middle of the
slanted interface. If there is a clear layer of yellow liquid on top of the
cream include it as part of the cream layer.
F. Press the ENT button. The LCD will display the creamatocrit result.
ENT
CRCT = XX.X%
Cal/oz MENU
RUN
G. Press ENT to display calories per ounce. Press RUN to return to the main
menu.
ENT
Cal = XX.X/oz
kcal/100 ml MENU
RUN
H. Press ENT to display kcalories per 100ml. Press RUN to return to the
main menu.
ENT
kcal/100ml = XXX.X
kcal/L MENU
RUN
I. Press ENT to display kilocalories per liter. Press RUN to return to the
main menu.
ENT
kcal/L = XXXX.X
Fat g/L MENU
RUN