Product Manual

15
SECTION 10
SPECIFICATIONS
Use: For the centrifugation of breast milk
Part Numbers: 100-146, 100-158
Reading Range: The effective creamatocrit reading range is 1.1% to 50.9%
Rotor: 6 place fixed angle head
Capacity: Six – 75mm by 1.1 mm I.D. Creamatocrit tubes
Max Rotor
Capacity: Six - 75mm length, 1.1 ID Creamatocrit Plus tubes filled ¾ full
Tube Holders: Transparent plastic disposable - 10 supplied in plastic bag
Motor: Sealed, ironless core, permanent magnet, DC
Lid: Safety interlocked to prevent opening while rotor is spinning and
includes a lid gasket
Timer: Fixed at three minutes. Timer operates for 180 seconds when
the spin cycle is engaged. A series of BEEPS signals when
rotor stops.
Front Panel: LCD prompts operational modes and alarms. Two tactile
membrane switch buttons engage various functions and options.
Speed: Designed to operate between 5,670 - 6,930 rpm and 1,548 –
2,312 RCF
Ramp-Up Time: Less than 10 seconds
Power Supply: Input Volts 100-240VAC± 10%, 0.7 Amps Frequency 47-63Hz
Output Volts 20 VDC, 1.2 Amps
Optional
Battery: Rechargeable nickel metal hydride, 12 cells, 1550 maH, 14.4 V
nominal
Dimensions: W. 7", L. 11", H. 5"
Bench Area
Requirements: W. 9", L. 13", H. 12"
Conditions: For indoor use only. Maximum relative humidity 95%.
Non-condensing.
Temperature
Range: Operating: 10ºC to 40ºC 50ºF to 104ºF
Storage: -10ºC to 60ºC 14ºF to 140ºF
Weight: Centrifuge 2 lbs.; Total shipping weight with battery 8 lbs.
Approvals: ETL/CETL Listing for North America
UL 61010-1
CAN/CSAC22.2#61010-1
IEC 61010-1
CENELEC EN 61010-1
IEC 61010-2-020