Product Manual

8
rpm. Once achieved, the spin time will be displayed on the LCD. Start
timing the spin when WAIT 180 SEC is displayed on the LCD. Stop timing
the spin when the motor shuts off.
B. RPM
The Creamatocrit Plus is designed to operate between 5,670 and 6,930
rpm. An internal microprocessor continuously monitors the rpm during
operation. If the rpm should drop below the specified range, the motor will
shut off, and the message LOW RPM will display on the LCD. To clear the
LOW RPM message, press the ENT or RUN button to return to the main
menu. Press the RUN button to restart the spin cycle. If the LOW RPM
message is displayed again, refer to Section 6. The rpm reading on the
LCD should be within 2% of a tachometer reading.
SECTION 5
MAINTENANCE
5.1 CLEANING
As with all electrical devices, make sure the centrifuge is unplugged before
cleaning. Always wear protective clothing when using any cleaning materials.
NEVER USE BLEACH, ABRASIVES OR CORROSIVE SOLVENTS.
DO NOT SPRAY OR ALLOW ANY LIQUID TO GET INSIDE THE
CENTRIFUGE. LIQUID WILL HARM THE ELECTRONICS. SUBSEQUENT
PROBLEMS WILL NOT BE COVERED UNDER WARRANTY.
BEFORE USING ANY CLEANING OR DECONTAMINATION METHODS
EXCEPT THOSE RECOMMENDED BY THE MANUFACTURER, USERS
SHOULD CHECK WITH THE MANUFACTURER THAT THE PROPOSED
METHOD WILL NOT DAMAGE THE EQUIPMENT.
Use a disinfectant towelette or a cloth slightly dampened with any non-corrosive
disinfectant solution to clean the lid and other parts of the centrifuge housing.
Dry all surfaces with a soft tissue or cloth after cleaning.
The rotor should be removed and cleaned regularly. Remove the rotor from the
motor shaft by first unscrewing the rotor knob. Gently lift the rotor vertically off
of the motor shaft. Make sure the rotor is thoroughly dry before reinstalling.
Liquid left on the rotor will cause damage to the device. Re-install the rotor
making certain that the rotor knob is tight.
If hazardous material is spilled on or inside the centrifuge, the user should carry
out appropriate decontamination.