Product Manual
6
J. Press ENT to display fat in grams per liter. Press RUN to return to the
main menu.
ENT
Fat = XXX.X g/L
CRCT MENU
RUN
K. Press ENT to display CRCT. Press RUN to return to the main menu.
3.6 CREAMATOCRIT DETERMINATION REMINDERS
TO ENSURE CORRECT RESULTS:
Use only Creamatocrit Plus tubes.
Spin sample one time (3 minutes) ONLY.
Read the tube immediately after removing it from the rotor.
Place the vertical black line of the slider in the middle of the MILK / CREAM
and CREAM / AIR full slanted interfaces.
Occasionally you may see a “tail” or line of cream running down the inside of
the tube. It will not affect the result. Simply ignore it when marking the
interfaces on the tube.
Do not smooth the holes out of the sealant in the tray with your finger. This
can result in sealant blowouts.
Tubes spun in a Creamatocrit Plus should be read on the Creamatocrit Plus.
Inspect and replace dirty or cracked disposable plastic tube holders monthly
or more frequently if necessary.
3.7 CENTRIFUGE REMINDERS
Do not lean on the centrifuge
Do not stay in the centrifuge envelope longer than is necessary for
operational reasons.
Do not deposit potentially hazardous material within the clearance envelope.
Do not use flammable or explosive materials in the centrifuge.