User`s manual
RadiCS
Chapter 10 Information
87
English
Correlation Between IEC, JIS and RadiCS
Pattern Check
Screens without the aspect ratio 1:1 are not applicable since the aspect ratio of the test pattern in IEC
and JIS is 1:1. Therefore, RadiCS has patterns with its own solution for each.
Test Pattern 1 The same as Bild 2 pattern. Scaling must be in accordance with a screen
resolution.
Test Pattern 2 The same as Bild 1 pattern. Scaling must be in accordance with a screen
resolution.
Test Pattern 3 The same as Bild 4 pattern. Scaling must be in accordance with a screen
resolution.
Test Pattern 4 The same as Bild 5 pattern. Scaling must be in accordance with a screen
resolution. Delete Test Pattern 4 if possible because it is not required on
monochrome monitors.
Luminance Check
Δ
shows the deviation value by comparing it with the basic value. For instance,
Δ
L
’
max < 20% means
the deviation value
±
20%. It compares the L
’
max value with the basic value.
The following values will be input in the basic values as initial setup values when performing the
tasksetup.
Δ
L’max Lmax value for calibration setup
Δ
L’min Lmin value for calibration setup or Lmin value the monitor recommended.
Δ
Lamb 0.01
It takes at least one week to establish a basic value. Therefore, monitors need to be tested
continuously. The average of a measurement result is a baseline value in luminance.
Sensors
LS-100, CD-Lux, LX-Plus and LX-Can are available measuring devices for RadiCS.
EIZO Sensor is available for constancy tests.
Cautions
The geometrical distortion of RadiForce series monitors does not need to be measured because it
meets the requested specification. However, non-RadiForce monitors may be used. Therefore, the
pattern check has patterns and checkpoints for geometrical distortion.