Chapter 8 Other Functions
64
English
Setting
Note
This operation can only be carried out in the advanced mode.
M
onitor status monitoring function can be used with a monitor of EIZO RadiForce series and
FlexScan MX series (except for some models).
F
or using Backlight Meter, conform that the monitor mode is set to be calibrated, and then set the L
max as a default (the value displayed by clicking [Default] in the “Calibration Settings” window) (see
page 19 “3-1. Calibration with an Optical Sensor”).
[Procedure]
1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting
dialog appears.
2. Clic
k [Monitor Status] to display the monitor status screen.
Status Analyzer
Enable Status Analyzer Diagnoses the monitor 35 minutes after logon.
Backlight Meter
Enable BackLight Meter To estimate the lifetime, performs the backlight
monitoring and displays the monitor status 35
minutes after logon.
Threshold Enter the threshold to display the backlight status.
Input range: 5 - 50
3. One of the following icons is displayed on the task tray.
OK: Normal
Warning: Status Analyzer is enabled, and there is a monitor with “Failed”
diagnostics result at logon
Caution: Backlight Meter is enabled, and there is a monitor that Backlight
Meter status is “Caution”
Caution and
Warning:
The following conditions are met
Status Analyzer is enabled, and there is a monitor with “Failed”
diagnostics result
Backlight Meter is enabled, and there is a monitor that
Backlight Meter status is “Caution”