Chapter 5 Monitor Setup
45
English
5-3. Diagnostics
Judges if luminance is managed properly with monitor luminance information.
The calibration setting must be changed and the message to re-calibrate a monitor
disappears if the luminance happens to be lower as a result.
[Procedure]
1. A monitor m
ust be on 30 minutes before diagnosing.
2. Click [Diagnostics] on the menu to diagnose.
3. A message of the diagnostics result will appear. Follow the directions.
Note
• Diagnose a monitor when it becomes stable (30 minutes later than the input signal is
switched after turning the power on). An error may appear if monitor display is not
stable.
• Exit
the user mode, change the calibration target in the advanced mode and calibrate the
monitor if the monitor luminance is lower.
5-4. Manual Measurement
This function manually measures luminance.
[Procedure]
1. Clic
k “Manual Measurement” on the menu.
2. Place the sensor in the measurement position and click the [Measure] button.
Note
• Measurements cannot be made using the Swing Sensor.