User`s manual
RadiCS
Chapter 4 Acceptance Test / Constancy Test
40
English
6. To proceed to the pattern check, follow the directions.
Tips
• When executing the pattern check for multiple monitors, the “Maker Model S/N: Serial
only” check box is displayed. To judge all monitors at the same time, disable the check
box. To judge all monitors separately, enable the check box.
• T
o display the pattern to full screen, click the left mouse button on the pattern with a
displayed message. To re-display a message, click the left mouse button again.
• A
pattern check message can be omitted. Click the right mouse button on the pattern
and click “Show check points” to delete a check mark. When check points are not
displayed, click the right mouse button and click [OK] on the menu, or press “Enter” on
the keyboard. It will judge that the displayed pattern check result has no problem and
will display the next pattern.
7. Click [Yes] if the questions in Check Point are correct. If there are any problems,
click [No].
Tips
• Enable the [Comment] check box to enter some comments of the check result in the
comment dialogue. Enter the desired comments (within 450 characters) when the
Comment dialogue box is displayed. A question mark “?” appears in the history
judgment when some comments are entered.
8. Click [Retry] to perform the pattern check again or click [Exit] to exit the check.
To execute the pattern check for other monitor, click [Next Monitor].
Luminance Check
Checks black and white luminance.
[Procedure]
1. Clic
k [Task] - [Luminance check].
2. “Input tester name” will appear. Input tester’s name and click [OK].
Note
• Input tester’s name in 31 characters.
3. Click [Proceed].
Tips
• Click [Skip] to skip a monitor luminance check onto the next monitor.
4. Follow the directions of the luminance check.
5. Click [Next Monitor] to check luminance of a different monitor. Click [Exit] to exit
the check.
Tips
• For GS521-ST, if Luminance Check has not been executed, a black image is displayed in full screen.