User`s manual
RadiCS
Chapter 4 Acceptance Test / Constancy Test
34
English
Sensor Setting
[Procedure]
1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting
dialog appears. Click “Sensor” and specify a sensor for acceptance test.
2.
To use a EIZO sensor, set up monitor ambient luminance input.
See “5-1. Ambient Luminance Input”.
Note
• Set up LS-100, CD-Lux, CD mon, LX-Plus or LX-Can as a sensor if selecting DIN or
QS-RL for the acceptance test standard. An acceptance test cannot be performed with an
EIZO sensor.
Executing Acceptance Test
[Procedure]
1. To execute the acceptance test, click “Proceed”.
Tips
• Acceptance test can also be executed by clicking [Task] - [Acceptance test] on the menu
bar.
2. “Input tester name” appears. Input the tester’s name and click [OK].
Note
• Input tester’s name in 31 characters.
3. Click [Proceed].
Tips
• To skip the current monitor pattern check and move onto the next monitor, click [Skip].
4. To proceed to the acceptance test, follow the directions on the message.
5. To execute the acceptance test of other monitors, click [Next Monitor]. To exit the
acceptance test, click [Exit].
Tips
• If the following conditions are met, the message to confirm the change of the baseline value appears
after the luminance check.
- The monitor quality control standard of the acceptance test is DIN or QS-RL
- The monitor quality control standard of the constancy test is the same one as the acceptance test
Click [Yes] to set each baseline value for the constancy test with the L’max, L’min and Lanb
measured in the acceptance test, click [No] when you does not set the value.