Chapter 4 Acceptance Test / Constancy Test
33
English
Monitor Luminance (Lamb)
Judgment Select a judgment standard from the pulldown menu.
Enter the baseline value between 0.00 to 9.99 if selecting Delta
Lamb < 25% or < 30%.
Multi-Monitor
Delta L’max (%) Enter the maximum allowable difference as a percent ratio when
comparing the Lmax of each monitor. (Setting range: 0 - 100).
Delta L’min (%) Enter the maximum allowable difference as a percent ratio when
comparing the Lmin of each monitor. (Setting range: 0 - 100).
Delta (L’max/L’min)
(%)
Enter the maximum allowable difference as a percent ratio
when comparing the Lmax/Lmin of each monitor.
(Setting range: 0 - 100).
Tips
Compares same model monitors.
4. Open [Grayscale] and set up the error check.
Execute Error
Check
The check box must be valid when the grayscale is scanned.
Target Error Rate
(%)
Enter 0 to 100 in the error rate when the ratio to the
measurement value (cd/m²) is calculated.
And enter 3 to 256 in the screen measurement point.
And then enter 0 - 100 in the error rate when the ratio to GSDF
is calculated.
5. Open uniformity and set up the measurement standard.
Number of
Measurement
Point
Select a measurement point of the screen.
Window Size (%) Set up the measurement window size for 5 - 50.
Judgement
(Luminance
Uniformity)
Set up the error judgement standard of luminance uniformity. An
error judgement standard can be set for 2 grayscale uniformities.
The check box must be valid when it is scanned.
Judgement
(Color Uniformity)
Set up the error judgement standard of white uniformity.
6. Open “Task” and enable the check box for acceptance test.
7. Click [OK] to close the acceptance test window and click [OK] on the monitor
when all the tab setups are completed.