Cover Important Please read this User’s Manual carefully to familiarize yourself with safe and effective usage.
Copyright © 2004-2010 EIZO NANAO CORPORATION All rights reserved. No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electric, mechanical, or otherwise, without the prior written permission of EIZO NANAO CORPORATION. EIZO NANAO CORPORATION is under no obligation to hold any submitted material or information confidential unless prior arrangements are made pursuant to EIZO NANAO CORPORATION's receipt of said information.
RadiCS Table of Contents Table of Contents Cover - - 1 Table of Contents - - 3 Chapter 1 Introduction - - 8 1-1. Features - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8 1-2. RadiCS Overview- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8 1-3.
RadiCS Table of Contents 4-2.
RadiCS Table of Contents Chapter 8 Other Functions - - 61 8-1. Icons in Notification Area- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 61 8-2.
RadiCS Table of Contents Chapter 10 Information - - 77 1. Standards- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 77 1-1. Quality Control Standards for Medical Imaging Display Monitors (Monitor Quality Control Standards)- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 77 AAPM On-line Report No.
RadiCS Table of Contents IEC, JIS - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 86 RadiCS Setup- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Correlation Between IEC, JIS and RadiCS - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Pattern Check- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Luminance Ch
RadiCS Chapter 1 Introduction Chapter 1 Introduction RadiCS organizes the image quality managements and the record of it. 1-1.
RadiCS Chapter 1 Introduction 1-3. Software Structure RadiCS consists of “User Mode” and “Advanced Mode”. Operators check on monitors everyday. An administrator sets up applicable standards and check items. Both of them use the same application software by switching the operation mode. User Mode Makes it possible to check operations required for Monitor quality control. Just simply follow the wizard to check some functions without any complicated setups.
RadiCS Chapter 2 RadiCS Setup Chapter 2 RadiCS Setup 2-1. System Requirement Here’s What You Need to Install RadiCS Operating System • Windows XP Professional Service Pack2/Service Pack3 Windows XP Professional x64 Edition Service Pack2 - PC with 1 GHz or 64-bit (x64) processor clock speed required. - 256 MB of RAM or more required. - Microsoft Internet Explorer 6.0 Service Pack2 Microsoft Internet Explorer 7.0 Microsoft Internet Explorer 8.
RadiCS 11 Chapter 2 RadiCS Setup Compatible Sensors • EIZO UX1 sensor • LX-Can • EIZO Clip-On Swing Sensor G1 • LS-100 (excluding GS521-ST) • EIZO Clip-On Swing Sensor G2 (excluding GS521-ST) • CD-Lux (firmware Ver.1.95 or later supported) • CD mon • EIZO Built-in Swing Sensor • EIZO Integrated Front Sensor • LX-Plus • MAVO-SPOT 2 USB Note • The available functions vary depending on the sensor used. Monitor • For monitor compatibility, please refer to our website. 2-2.
RadiCS Chapter 2 RadiCS Setup 2-3. Installation Procedure [Procedure] 1. Insert CD-ROM into a CD-ROM readable drive. The menu starts up. Tips • Double-click “Launcher.exe” in CD-ROM when the menu does not start up automatically. 2. Click “Install RadiCS” and follow the software directions to install it. 3. When the “Set Password” screen appears, specify the password that will be used to start the RadiCS software in the advanced mode.
RadiCS 13 Chapter 2 RadiCS Setup 2-4. Uninstallation Procedure Follow the procedure below to uninstall RadiCS software. Windows XP [Procedure] 1. Open [Control Panel] - [Add and remove programs]. Select “RadiCS” and “RadiCS Network Upgrade Software”. 2. Click [Remove] and follow the directions to uninstall RadiCS. Windows Vista / Windows 7 [Procedure] 1. Open [Control Panel] - [Programs] and click [Uninstall a Program]. Select “RadiCS” and “RadiCS Network Upgrade Software”. 2.
RadiCS Chapter 2 RadiCS Setup 2-5. How to Get Ready for an Operation Start up and set up RadiCS in the advanced mode after installing its software. Procedure Set up a monitor Display a main screen According to the wizard, •Select the screen display state •Input monitor information Startup [Procedure] 1. Click [Start] - [All Programs] - [EIZO] - [RadiCS] (Advanced Mode). 2. Enter the password you specified during the installation and then click [OK]. 3.
RadiCS 15 Chapter 2 RadiCS Setup [Procedure] 1. A circle is displayed on the monitor being set, so select how the circle is shown on the screens. 2. When Wide View or Mirroring is selected, the following setup screen is displayed. Wide View Specify the number of screens and the configuration. Mirroring Specify the number of screens.
RadiCS Chapter 2 RadiCS Setup 3. Register the monitor information. When using an EIZO monitor (except for some models) Click [Select Monitor] to display a list of monitors connected by USB connection. Select the monitor and click [OK]. The Monitor Type, Manufacturer, Model Name and Serial Number (S/N) are automatically entered. Set the other items as necessary. Note For GS521-ST: ・Monitor information is registered/managed as 2 monitors.
RadiCS 17 Chapter 2 RadiCS Setup Monitor Information Window Register this monitor Disable the check box when not controlling the monitor. Monitor Type Select the monitor type from Color LCD, Monochrome LCD, Color CRT, or Monochrome CRT. Select Monitor Displays the “Select Monitor” dialogue. When you select a monitor from the list, its information is automatically entered. Manufacturer/ Model Name/ Serial Number (S/N) Enter the manufacturer, model name, and serial number (S/N).
RadiCS Chapter 2 RadiCS Setup 6. To correlate data during the monitor measurement, click [Correlation]. Tips • When RadiCS detects two or more monitors to which Clip-On Swing Sensor is connected, the [Correlation] check box appears. When the check box is enabled, multiple monitors can be correlated simultaneously (the check box does not appear when only one monitor is detected). 7. Attach an EIZO sensor (UX1) in the center of the measurement window, and click [Proceed]. Correlation starts.
RadiCS 19 Chapter 3 Monitor Calibration Chapter 3 Monitor Calibration The monitor calibrates itself regularly to keep the screen display constant. And a monitor also calibrates itself when another monitor is set up or monitors are changed. The monitor calibrates itself with either an optical sensor or a backlight sensor. Note •This operation can only be carried out in advanced mode. Optical Sensor: Measuring Standard, Target, Display Function Backlight Sensor: 7500K or 12500K 3-1.
RadiCS Chapter 3 Monitor Calibration 4. Click the right mouse button on the monitor and select “Calibration setting”. 5. Set up each item. Tips • Click [Default] to set the default values. Note • When calibrating an EIZO FlexScan M series monitor or a monitor made by another manufacturer using a LS-100 sensor, disable the “Lamb” check box. Measurement Level Low Standard High Target Lmax Lmin Color Shorten the measuring time. Set up the default. Emphasize the measurement accuracy. It takes longer.
RadiCS 21 Chapter 3 Monitor Calibration Display Function DICOM Part14 GSDF CIE Exp Log Linear Linear User Definable The setup complies with DICOM Part14. The monitor luminance is set up being reflected from the environment light when calibrating. It will also be reflected from environment light when checking luminance and grayscale. Reference: Lmax+Lamb = Maximum Luminance Target Lmin+Lamb = Minimum Luminance Target Enter from 0.00 to 9.99 to set up the monitor luminance.
RadiCS Chapter 3 Monitor Calibration How to Save the Calibration Setting in a File Several calibration settings are available after a calibration setting is saved in a file and saved in an optional folder. [Procedure] 1. Click [File] - [Save as] after changing calibration settings. 2. Specify a place, name the file and click [OK]. Click [File] - [Open] to open the file when the saved file needs to be read into the calibration setting window. Sensor Setting [Procedure] 1.
RadiCS 23 Chapter 3 Monitor Calibration Swing Sensor/Integrated Front Sensor [Procedure] 1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting dialog appears. Click [Monitor List] to display the monitor list screen. Currently connected monitors appear in the list, grouped by the type of graphics board. 2. Locate the monitor to which the Swing Sensor/Integrated Front Sensor is installed, select Sensor from the tree, and click the right mouse button.
RadiCS Chapter 3 Monitor Calibration *Detection level setting Level Detectable Range Clip-On Swing Integrated Sensor G2 Front Sensor Level-1 70 cm or less 120 cm or less Details When you get out of detectable range, the schedule is executed. This setting is recommended if the schedule is often executed Level-2 90 cm or less 90 cm or less when the level is set to “Level-3” or “Level-4” even though you are in front of the monitor.
RadiCS 25 Chapter 3 Monitor Calibration How to Calibrate a Monitor Tips • Refer to page 27 when calibrating an EIZO FlexScan M series monitor or a monitor made by another manufacturer. [Procedure] 1. Turn the monitor on 30 minutes before calibration. 2. Attach an EIZO sensor (UX1). 3. Click [Proceed] in calibration. The monitor can also be calibrated by clicking [Calibration] - [Optical sensor] on the menu bar. 4. [Input tester name] will appear. Input tester’s name and then click [OK].
RadiCS Chapter 3 Monitor Calibration 6. Attach the sensor to the measurement window then click [Proceed]. Tips • A calibration message and a measurement window are displayed in each monitor when several monitors are connected. Click [Skip] to move a message onto the next monitor when a message and a measurement window are displayed on the monitor. 7. Follow the instructions on the message to continue calibration. 8. The result will appear when finishing calibrating a monitor connected to the computer.
RadiCS 27 Chapter 3 Monitor Calibration When using an EIZO FlexScan M series monitor or a monitor made by another manufacturer Calibration is available when using an EIZO recommended graphics card and the “Independent” monitor setup. Note When using a LS-100 sensor, confirm that the settings are as follows: • “Monochrome LCD” is selected as a monitor type in the Monitor Setup window regardless of the monitor type (color/monochrome) (see page 16 “When using an another manufacturer’s monitor”).
RadiCS Chapter 3 Monitor Calibration 7. The Lmax adjustment window is displayed. Click [Calibration settings] and then set the target (Lmax:). (1) (2) 8. Click [Measure]. The luminance is measured by the sensor and the measurement value is displayed. Use the monitor brightness adjustment function to set the luminance to the (1) target displayed in the Lmax adjustment window. 9. Click [Measure]. The [OK] button becomes valid when the measurement value is the (2) target or higher. 10.Click [OK].
RadiCS 29 Chapter 3 Monitor Calibration 3-2. Calibration with a Backlight Sensor Monitors are calibrated and color-adjusted with a backlight sensor. Note • IThe color temperature and the target value of calibration setting must be set 7,500K or 12,500K (for MX190S, 6,500K or 12,500) when proceeding this function.* (See “3-1. Calibration with an Optical Sensor”.) • The color adjustment is available when the monitor is R22. [Procedure] 1. The monitor must be on 30 minutes before proceeding calibration. 2.
RadiCS Chapter 4 Acceptance Test / Constancy Test Chapter 4 Acceptance Test / Constancy Test 4-1. Acceptance Test Test acceptance of RadiCS before operations when monitors are set up or changed. The acceptance test standards include the following nine types: DIN, AAPM, ACR, JESRA, QS-RL, EUREF, IPEM, BasicQC and DMG QC Manual. The test items can be customized depending on requirements. Note • This operation can only be carried out in the advanced mode. Select a sensor.
RadiCS 31 Chapter 4 Acceptance Test / Constancy Test Description Pattern List (No, Reversed, Pattern) Pattern Selection Add Delete Move Up Move Down Preview Bitmap Image Default Check Point Description* Edit the descriptive text for a pattern set. (See “How to Edit the Description”) A list of patterns for the pattern check appears. “Reversing Display” allows the display in reverse black and white order (See “Chapter 9 Patterns”).
RadiCS Chapter 4 Acceptance Test / Constancy Test How to Edit the Description Enter the descriptive text of up to 128 one-byte characters in “Description”. How to Edit a Check Point Enter 450 characters or less into the checkpoint. Note • Users must answer “YES” if there is no problem and must answer “NO” if there is a problem. Please consider the following things when making questions. - Questions must be in question formation. e.g.
RadiCS 33 Chapter 4 Acceptance Test / Constancy Test Monitor Luminance (Lamb) Judgment Select a judgment standard from the pulldown menu. Enter the baseline value between 0.00 to 9.99 if selecting Delta Lamb < 25% or < 30%. Multi-Monitor Delta L’max (%) Delta L’min (%) Delta (L’max/L’min) (%) Enter the maximum allowable difference as a percent ratio when comparing the Lmax of each monitor. (Setting range: 0 - 100).
RadiCS Chapter 4 Acceptance Test / Constancy Test Sensor Setting [Procedure] 1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting dialog appears. Click “Sensor” and specify a sensor for acceptance test. 2. To use a EIZO sensor, set up monitor ambient luminance input. See “5-1. Ambient Luminance Input”. Note • Set up LS-100, CD-Lux, CD mon, LX-Plus or LX-Can as a sensor if selecting DIN or QS-RL for the acceptance test standard.
RadiCS 35 Chapter 4 Acceptance Test / Constancy Test 4-2. Constancy Test Monitors for diagnostic imaging use must be tested regularly upon display performance. (= constancy test) A constancy test has pattern/luminance/gray scale/uniformity. The check items and judgment standards vary depending on a constancy standard. The constancy test includes eleven different standards: DIN, AAPM, ACR, IEC, JIS, JESRA, QS-RL, EUREF, IPEM, Basic QC and DMG QC Manual.
RadiCS Chapter 4 Acceptance Test / Constancy Test 2. Click “Task” and enable the check box of the item to whicn the constancy test is executed. 3. Click “Pattern” and set up a pattern for the constancy test. The contents set in the task setup wizard appear and can be confirmed. Set 1 - 5 Description Allows up to five types of pattern sets to be set. Edit the descriptive text for a pattern set. (See “How to Edit the Description”) Pattern List A list of patterns for the pattern check appears.
RadiCS 37 Chapter 4 Acceptance Test / Constancy Test How to Display a Pattern for User Setting 1) Select [User...] from “Pattern Selection” pulldown menu and open “User Setting” window. 2) Click [Add Pattern], select an optional test pattern file and click [Open]. A pattern will be added to the pattern list. 3) Select a pattern to add to the pattern list and click [OK] to return to “Pattern”. 4) Enter a question in the pattern check if necessary. 4. Open [Luminance] and set up a luminance test standard.
RadiCS Chapter 4 Acceptance Test / Constancy Test 5. Open [Grayscale] and set up the error check. Execute Error Check Target Error Rate (%) The check box must be valid when grayscale is scanned. • Enter 0 to 100 in the error rate when the ratio to the screen measurement value (cd/m²) is calculated. • Enter 3 to 256 in the screen measurement point. Enter 0-100 in the error rate when the ratio to GSDF is calculated.
RadiCS 39 Chapter 4 Acceptance Test / Constancy Test Executing Constancy Test Pattern Check Displays test patterns and allows an operator to see if specific pattern classification is possible with the human eye. For the monitors with the following conditions, the simultaneous pattern check for multiple monitors is available. • Monitors that are made by the same manufacturer and that are the same model. • EIZO monitors with CAL Switch function that are the same model and have the same CAL mode.
RadiCS Chapter 4 Acceptance Test / Constancy Test 6. To proceed to the pattern check, follow the directions. Tips • When executing the pattern check for multiple monitors, the “Maker Model S/N: Serial only” check box is displayed. To judge all monitors at the same time, disable the check box. To judge all monitors separately, enable the check box. • To display the pattern to full screen, click the left mouse button on the pattern with a displayed message.
RadiCS 41 Chapter 4 Acceptance Test / Constancy Test Grayscale Check Checks grayscale. [Procedure] 1. Click [Task] - [Grayscale check]. 2. “Input tester name” will appear. Input tester’s name and click [OK]. Note • Input tester’s name in 31 characters. 3. Click [Proceed]. Tips • Click [Skip] to skip a monitor grayscale check onto the next monitor. 4. Follow the directions of the grayscale check. 5. Click [Next Monitor] to check grayscale of some other monitor. Click [Exit] to exit the check.
RadiCS Chapter 5 Monitor Setup Chapter 5 Monitor Setup The advanced mode has some setups and functions for RadiCS software besides acceptance test setting and calibration setting. Each connected monitor can be set on its own. [Procedure] 1. When selecting [Configuration](Advanced mode)/[Configuration/View](User mode) on the menu bar and then clicking [Change Settings], the setting dialog appears. Click [Monitor List] to display the monitor list screen.
RadiCS 43 Chapter 5 Monitor Setup Pattern Diagnostics Manual Measurement Identification Property Configuration* View A pattern is displayed at the top of monitor Images (an optional monitor or all connected monitors) See 5-2. Obtain monitor luminance information and judge if the current luminance is managed right. See 5-3. Manually measures luminance. See 5-4. Display a manufacturer, a model and a serial number on a selected monitor for 3 seconds. See 5-5.
RadiCS Chapter 5 Monitor Setup 5-2. Pattern Display Displays a pattern image used for acceptance test/constancy test on either an optional monitor or all connected monitors. It displays only a selected pattern and does not set up or check patterns. Tips • See Chapter 9 for description of each pattern. [Procedure] 1. Click “Pattern” on the menu. 2. Select a pattern and click [Display]. To display a pattern for user setting, click [User...], select a pattern and click [OK].
RadiCS 45 Chapter 5 Monitor Setup 5-3. Diagnostics Judges if luminance is managed properly with monitor luminance information. The calibration setting must be changed and the message to re-calibrate a monitor disappears if the luminance happens to be lower as a result. [Procedure] 1. A monitor must be on 30 minutes before diagnosing. 2. Click [Diagnostics] on the menu to diagnose. 3. A message of the diagnostics result will appear. Follow the directions.
RadiCS Chapter 5 Monitor Setup 5-5. Identification Displays the identification information of each monitor. [Procedure] 1. Click “Identification” on the menu. 2. “Manufacture/Model Name/Serial Number (S/N)” will appear for 3 sec. 5-6. Property Displays information set in the monitor setup. [Procedure] 1. Click “Property” on the menu. 2. Monitor information will appear. Tips • To edit information displayed in the property, run Monitor Setup in the advanced mode.
RadiCS 47 Chapter 6 Scheduling Chapter 6 Scheduling 6-1. How to Set Up RadiCS Schedules Tasks for constancy test, diagnostics, and calibration (checks of patterns, luminance, gray scale, uniformity) can be performed regularly by setting schedules. Calibration schedules can also be set in the case that Swing Sensor/Integrated Front Sensor is being used. Note • This operation can only be carried out in the advanced mode. [Procedure] 1.
RadiCS Chapter 6 Scheduling Activate Scheduler Enable the check box to use a scheduler. Enable the corresponding check box to execute Pattern Check, Luminance Check, Grayscale Check, Uniformity Check, Calibration, Calibration (Backlight Sensor), or Diagnostics.
RadiCS 49 Chapter 6 Scheduling 3. Click [OK] when the setting is completed. Tips • A scheduler icon appears on the Windows task tray when a scheduler starts up. Windows group policy settings To execute Monitor Quality Control during shutdown, following the configurations need to be set. Note • If “Proceeds the task before shutdown” is enabled for a schedule, this configuration must be set. If the configuration is not set correctly, Monitor Quality Control will not be executed.
RadiCS Chapter 6 Scheduling 4. On the “Group Policy” screen, click [Computer Configuration] - [Administrative Templates] - [System] - [Scripts]. Right-click [Maximum wait time for Group Policy scripts] and select [Properties]. 5. The “Maximum wait time for Group Policy scripts Properties” screen appears. Select “Enabled”, enter “32000” (recommended value) into “Seconds:”, and click [OK].
RadiCS 51 Chapter 6 Scheduling peration when Executing a Scheduler Task The message below appears when a task is executed by a scheduler for the following checks: • Pattern check • Luminance check, grayscale check, or uniformity check using the EIZO Sensor, LX-Plus, LX-Can, CD-Lux, CD mon, LS-100 or Manual Input. Note • This message does not appear for the luminance check and grayscale check using the Swing Sensor. ex.) Pattern check Click [Yes] to display a tester’s name input screen.
RadiCS Chapter 6 Scheduling 6-3. How to Set Up SelfQC Schedules Schedules for calibration and grayscale check of the monitor with the RadiCS SelfQC* function can be set. *Calibration and grayscale check can be executed by only a monitor. Note • A SelfQC compatible monitor is required. • This operation can only be carried out in the advanced mode. [Procedure] 1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting dialog appears. 2.
RadiCS 53 Chapter 6 Scheduling Click [Change] in Task/Diagnostics to set up the followings Every day Every week Every month Quarter Biannual Annual Every login At application startup Set up a starting time. Set up a day and a starting time. Specify a date and a starting time. Specify a date, a month and a starting time. Sets the software so that pattern check is executed at logon. Sets the software so that pattern check is executed when the registered application is started. 3.
RadiCS Chapter 7 History Chapter 7 History Each monitor saves the record as history when acceptance and constancy are tested. The monitor image display quality can be managed based on the history and the history can be output as a report.
RadiCS 55 Chapter 7 History History Storage Folder History Reduction Reduce history Execute on Manage of Reduction Clear from the list Save as another file. Specify a history save folder. Enable the check box to reduce history. Specify a date monthly. Select a manner of reduction. Delete data when reducing history. Move data to a save folder when reducing history. Click [Reference] and specify an optional save folder when selecting this item.
RadiCS Chapter 7 History History List Date & Time Task CAL (Optical sensor)............calibration with an optical sensor CAL (Backlight sensor)........check with a backlight sensor CHK (Pattern)......................pattern check CHK (Luminance)................luminance check CHK (Grayscale)..................grayscale check CHK (Uniformity)..................uniformity check ToneCurve Tuning................ToneCurve Tuning Diagnostics .........................diagnostics LUT-CAL (Optical sensor)....
RadiCS 57 Chapter 7 History Tips • Click each item in the history list title to sort the record of the list according to clicked items. • Right-click each item in the history list title to set the item to be shown or hided. History Search Input conditions into the search condition area and click [Search]. Tips • A selected record in the history list is available as search conditions. Select a record and click the right mouse button to select “Set as a criteria”.
RadiCS Chapter 7 History Report Generator Generates an acceptance test and a constancy test report. In order to generate a report, select history data that meets the following conditions. Conditions for Generating a Report • The manufacturer of a monitor, the model and the serial number must match those of the monitor. • Test standard must be the same. • The required test results for each test must be complete. (If you customized test standard, it must be deleted from the conditions.
RadiCS 59 Chapter 7 History 7-3. Functions of Record Select a selected record in the history list and click the right mouse button to use the following functions. Advanced Mode User Mode Show details Displays detailed information of a selected record. Tips •Double-click a selected record in the history list window to display detailed information. • It is not available when the task of a selected record is diagnostics. Set as criteria See the history research.
RadiCS Chapter 7 History Export Exports selected recorded data into a file. Select “Export” to display a screen specifying where to export data to save. Retrieve LUT* Sets up to transfer stored adjustment data in the history to the monitor when a selected record is calibration or ToneCurve Tuning history. Tips • This function is not available when several history records are selected. Delete* Deletes selected records from the history list.
RadiCS 61 Chapter 8 Other Functions Chapter 8 Other Functions 8-1. Icons in Notification Area The following menu can be selected in the notification area. • RadiCS (User mode) • RadiCS (Advanced mode) • Pattern 8-2. Registration Information Setting Sets up for the information of a system adopting RadiCS software as registration information. Registered information will be used as a history function. Note • This operation can only be carried out in the advanced mode. [Procedure] 1.
RadiCS Chapter 8 Other Functions 2. Enter the necessary information and click [OK]. Registration Organization Address Phone Number Administration Location Department User Administrator Maintenance Service Provider Enter names of institutions within 64 characters. Enter addresses within 256 characters. Enter telephone numbers within 20 characters. Enter monitor installation sites within 32 characters. Enter departments, where monitors are used, within 32 characters.
RadiCS 63 Chapter 8 Other Functions 8-3. Monitor Status Monitoring Function With the following two functions, the monitor status is monitored and the results are displayed. - Status Analyzer: Executes diagnostics and displays the results. - Backlight Meter: Performs the backlight monitoring and displays the estimated monitor lifetime (the remaining time that the recommended brightness can be maintained). Confirmation [Procedure] 1.
RadiCS Chapter 8 Other Functions Setting Note • This operation can only be carried out in the advanced mode. • Monitor status monitoring function can be used with a monitor of EIZO RadiForce series and FlexScan MX series (except for some models). • For using Backlight Meter, conform that the monitor mode is set to be calibrated, and then set the L’ max as a default (the value displayed by clicking [Default] in the “Calibration Settings” window) (see page 19 “3-1. Calibration with an Optical Sensor”).
RadiCS 65 Chapter 8 Other Functions 8-4. Version Information [Procedure] 1. Click [Configuration](Advanced mode)/[Configuration/View](User mode)[About] on the menu bar. 2. RadiCS version information and its compatible monitors are listed. 8-5. Password Change You must enter a password when you start RadiCS in the advanced mode. This is also required if you want to change the password specified during the installation of RadiCS. Note • This operation can only be carried out in the advanced mode.
RadiCS Chapter 8 Other Functions 8-6. Sensor Setup Function If using the CD-Lux, LS-100 or Manual Input, you must enter the serial number of the sensor. [Procedure] 1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting dialog appears. 2. Click [Sensor]. Tips • When selecting Manual Input, the Input Measurement Value window appears during execution of each task.
RadiCS 67 Chapter 9 Pattern Chapter 9 Pattern RadiCS has a pattern to check patterns. Confirm a checkpoint of each pattern and the check patterns. SMPTE No. Description Check Points (1) The patches that are 5% and 95% of the backgrounds are inside. If they are very visible. (2) Each of the 12 cell-grayscales can be identified easily. If the grayscales are displayed evenly spaced apart. (3) The 0 to 255 grayscales are displayed and its continuity is identifiable.
RadiCS Chapter 9 Pattern TG18-QC (Reversing Display Available) Supporting standard: AAPM/ACR/JESRA/EUREF/IPEM/Basic QC/DMG QCM No. (1) Description The patches that are 5% and 95% of the backgrounds are inside. Check Points If they are very visible. (2) Each of the 16 cell-grayscales can be identified easily. If the grayscale is displayed evenly spaced apart. If the patch with the grayscale ± 4 at each corner is visible.
RadiCS 69 Chapter 9 Pattern Test 1 (Reversing Display Available) Supporting standards: DIN/IEC/JIS/QS-RL Description The crosshatch and the circle of 255 grayscales are on the background of 128 grayscales. The boarder of the pattern has “@” on it. Checkpoint Geometrical Characteristics Color Artifact Stability Convergence Test 2 4 intersections of borders at the top, the bottom, and the sides outside must be able to be confirmed visibly.
RadiCS Chapter 9 Pattern Test 3 Supporting standards: DIN/IEC/JIS/QS-RL Description White (255 grayscales) and black (0 grayscale) horizontal/vertical lines are in the center and the corner on the background of 128 grayscales. 2 different types of the lines are in the center. The contrasts of these types are different. Checkpoint Location Resolution Contrast Resolution Visible Check of Line Structures Test 4 All grids at modulation 100% must be identified without a loupe.
RadiCS 71 Chapter 9 Pattern Crosshatch (Reversing Display Available) Description Vertical and horizontal lines of 255 grayscales are displayed on the background of 0 grayscale. Checkpoint Convergence and Geometrical Characteristics Horizontal Lines There must be no convergence displacement and no distortion of straight lines. (Reversing Display Available) Description Horizontal lines of 255 grayscales are displayed on the background of 0 grayscale.
RadiCS Chapter 9 Pattern The right half-moon pattern is assigned with a grayscale 2 levels higher than the square; and the left half-moon pattern, 2 levels lower. The upper left and lower right patches, 4 levels higher; and the upper right and lower left patches, 4 levels lower. This pattern is mainly used to check the grayscale characteristics, but it can also be used as a simple check of the viewing angle characteristics. Watch the pattern from different angles.
RadiCS 73 Chapter 9 Pattern TG18-AD Supporting standard: AAPM/ACR Description There are grids of 7x7 in the center on the background of 0 grayscale. Each grid has a vertical line of a certain grayscale at every dot. The cell in row 1 and column 1 is 1 grayscale. 2 grayscale in column 2, 3 grayscale in column 3...7 grayscale in column 7. The cell in row 1 and column 1 is 8 grayscales. The cell in row 6 and column 7 is 49 grayscales. This pattern is to judge visible grayscales.
RadiCS Chapter 9 Pattern Gray Description A daubing pattern of 128 grayscales. Checkpoint Bailing glare Artifact A circle in the middle of a small circle is visible. No artifact must affect a diagnostics. There must be no deficiency. There must be no light color and dark color. No return wire is away from the bottom. Clinical Image (JESRA) (Reversing Display Not Available) Supporting standard: JESRA Description 15 artificial nodules are embedded in the simple chest X-ray image.
RadiCS 75 Chapter 9 Pattern JIRA CHEST-QC (Reversing Display Not Available) Supporting standard: JESRA (1) (1) (3) (2) (2) (3) Description The JIRA standard clinical image and the judgment area of the TG18-QC pattern are combined, so that the evaluation can be performed for each day of usage on a single image. The image has been further reduced for each resolution and added the gray circles to indicate the judgement areas.
RadiCS Chapter 9 Pattern TG18-LPH89, 50, 10 Supporting standard: EUREF TG-LPH89 TG-LPH50 TG-LPH10 Description Line pairs consisting of two single-pixel-wide horizontal lines with different grayscales are formed, covering the entire screen. The higher grayscale is equal to the lower grayscale +12%, and there are three types of patterns. The (high grayscale : low grayscale) values of the patterns are LPH89 = (255 : 228), LPH50 = (143 : 128), and LPH29 = (29 : 26).
RadiCS 77 Chapter 10 Information Chapter 10 Information Chapter 10 covers the following items; • Notes according to the monitor quality control standards for RadiCS. • Cautions when setting up tests on RadiCS with each monitor quality control standard. 1. Standards 1-1. Quality Control Standards for Medical Imaging Display Monitors (Monitor Quality Control Standards) AAPM On-line Report No.
RadiCS Chapter 10 Information IEC 61223-2-5: 1994 “Evaluation and routine testing in medical imaging departments - Part2-5: Constancy tests - Image display devices” formulated by the International Electrotechnical Commission. Tips • “IEC” written on RadiCS means “IEC 61223-2-5”. IPEM Report 91: 2005 “Recommended Standards for the Routine Performance Testing of Diagnostic X-ray Imaging Systems” formulated by Institute of Physics and Engineering in Medicine in the UK.
RadiCS 79 Chapter 10 Information JESRA X-0093*A-2010: 2010 “Quality Assurance (QA) Guideline for Medical Imaging Display Systems” prepared by Japan Industries Association of Radiological Systems (JIRA). It was published in 2005 and revised in 2010. It specifies the acceptance and constancy tests. The shipment test reports provided by manufacturers can be used as a substitute for the acceptance test.
RadiCS Chapter 10 Information SMPTE RP133: 1991 “Specifications for Medical Diagnostic Imaging Test Pattern for Television Monitors and Hard-Copy Recording Cameras” proposed by Society of Motion Picture and Television Engineers in the US. Tips • “SMPTE” written on RadiCS means “Test patterns created in reference to SMPTE PR113 specifications”. Basic QC The setting specific to RadiCS used for monitor management that does not comply with standards or guidelines described above.
RadiCS 81 Chapter 10 Information 2. About RadiCS Software 2-1. Prerequisite RadiCS Software We have long developed monitors. With those skills, knowledge and measuring data, we have developed RadiCS for users of medical image diagnosis to manage the quality of monitors efficiently according to our interpretation of the quality control standard for each medical image monitor.
RadiCS Chapter 10 Information Correlation Between DIN and RadiCS Pattern Check Screens without the aspect ratio 1:1 are not applicable since the aspect ratio of the test pattern in DIN is 1:1. Therefore, RadiCS has patterns with its own solution for each. Test Pattern 1 Test Pattern 2 Test Pattern 3 The same as Bild 3 pattern. Scaling must be in accordance with a screen resolution. The same as Bild 2 pattern. Scaling must be in accordance with a screen resolution. The same as Bild 5 pattern.
RadiCS 83 Chapter 10 Information AAPM RadiCS Setup Pattern Check Luminance Check Grayscale Check Uniformity Multiple Monitors Pattern Check Primary Acceptance Test Black TG18-QC TG18-AD TG18-UN80 TG18-AFC TG18-CT White (reversed Black) L’max/L’min > 250 L’max > 170 cd/m2 ΔL’max < 10 % Standard Value xxx cd/m2 *3 Target Error rate < 10 % of GSDF Grayscale: 204, 26 *4 Grayscale: 204 Δ(u’, v’) < 0.
RadiCS Chapter 10 Information Correlation Between AAPM and RadiCS Pattern Check Screens without the aspect ratio 1:1 are not applicable since the aspect ratio of the test pattern in AAPM is 1:1. Therefore, RadiCS has patterns with its own solution for each one. TG18-QC TG-AD TG18-AFC TG18-CT TG-UN80 The same as each pattern. Scaling must be in accordance with a screen resolution. Grayscale 204 white patterns.
RadiCS 85 Chapter 10 Information Cautions AAPM constancy testing has 3 types, tests that monitor users perform daily, tests that medical physicists perform or QC (quality control) technologists perform under their instructions monthly / quarterly, and tests that medical physicists perform annually. RadiCS is mainly intended for constancy testing of the second type, but pattern checks can be performed for all three types of testing.
RadiCS Chapter 10 Information Grayscale Check ACR includes an equality sign but RadiCS doesn’t because the target error rate is < 10% of GSDF. This is a judgment condition for DICOM Part 14 GSDF. The number of grayscale measuring points is fixed at 18 and is unchangeable. The measurement result is 17 points because it is expressed as (JNDn + 1 – JNDn) / 2. Uniformity Check ACR includes only a notation about chromaticity uniformity of Δ(u’, v’).
RadiCS 87 Chapter 10 Information Correlation Between IEC, JIS and RadiCS Pattern Check Screens without the aspect ratio 1:1 are not applicable since the aspect ratio of the test pattern in IEC and JIS is 1:1. Therefore, RadiCS has patterns with its own solution for each. Test Pattern 1 Test Pattern 2 Test Pattern 3 Test Pattern 4 The same as Bild 2 pattern. Scaling must be in accordance with a screen resolution. The same as Bild 1 pattern. Scaling must be in accordance with a screen resolution.
RadiCS Chapter 10 Information QS-RL RadiCS Setup Pattern Check Luminance Check Grayscale Check Uniformity Check Multiple Monitors Acceptance Test Category A Category B Test Pattern 1 Test Pattern 1 Test Pattern 2 Test Pattern 2 Test Pattern 3 Test Pattern 3 L’max/L’min > 100 L’max/L’min > 40 L’max > 200 cd/m2 L’max > 120 cd/m2 Lamb < L’max/100 Lamb < L’max/40 Grayscale: 128 *1 *1 (Lcorner-Lcenter) /Lcenter x 100 < 15 % *2 (Lcorner-Lcenter)/Lcenter x 100 < 20 % Pattern Check Luminance Check Gr
RadiCS 89 Chapter 10 Information Luminance Check Lmax and Lmin specified in QS-RL include the ambient luminance and are the same as L’max and L’min used in RadiCS. QS-RL specifies Lmin Delta 1.0 cd/m2, but RadiCS includes no equality sign. Lamb stands for the ambient luminance and refers to the same value as “Ls” of DIN. The inequality Lmax/Ls > 100 (or 40) in the standard has been transformed into Ls > Lmax/100 (or 40). In QS-RL, the luminance is specified as |Delta Ls| ≤ 0.3 Ls.
RadiCS Chapter 10 Information Cautions Category A Mammo conforms to the PAS1054 mammography standard. This standard includes monitor resolution of 2000 × 2500 or above as a condition, but RadiCS makes no such determination.
RadiCS 91 Chapter 10 Information Correlation Between JESRA and RadiCS Pattern Check Since the test patterns used in JESRA have the aspect ratio of 1:1, they cannot be applied directly to monitors. Thus, RadiCS determines necessary test patterns based on the inspection to be performed and generates its own patterns corresponding to the monitor’s resolution. TG18-QC TG18-UN80 Clinical Image and JIRA CHEST-QC (JESRA) Equivalent to the pattern with the same name in the standard.
RadiCS Chapter 10 Information Uniformity Check Although JESRA includes an equality sign, each judgement condition in RadiCS does not. In the standard, the uniformity is checked with the TG18-UN80 patterns. However, since these patterns have the aspect ratio of 1:1, they are not applicable directly. RadiCS displays two windows (grayscale: 204) with the size of 10 % of the whole display area at the center and a corner of the screen. It then measures the luminance at the center of both windows.
RadiCS 93 Chapter 10 Information Uniformity Check IPEM recommends using TG18-QC or SMPTE patterns, but these patterns are not suitable for measuring 50% grayscale uniformity. Instead, RadiCS displays grayscale 128 windows equivalent to 10% of the display area in the center of the screen and in the corners, and measures the center portion of each window. IPEM includes an equality sign but RadiCS does not.
RadiCS Chapter 10 Information Correlation Between EUREF and RadiCS Pattern Check The patterns used for EUREF are the same as those used for AAPM. RadiCS determines the properties to be verified and independently prepares appropriate patterns for each resolution. TG18-QC TG18-LPH (89,50,10) TG18-LPV (89,50,10) This is scaled to match the resolution. This is scaled to match the resolution. This is scaled to match the resolution.
RadiCS 95 Chapter 10 Information Basic QC RadiCS Setup Pattern Check Luminance Check Grayscale Check Uniformity Multiple Monitors TG18-QC Acceptance Test TG18-QC Constancy Test Correlation Between Basic QC and RadiCS Pattern Check The patterns used for Basic QC are the same as those used for AAPM. RadiCS determines the properties to be verified and independently prepares appropriate patterns for each resolution. TG18-QC This is scaled to match the resolution.
RadiCS Chapter 10 Information Luminance Check In DMG QCM, the luminance measurement does not include the ambient luminance. In RadiCS, an apostrophe (’) in the L’max, for example, indicates that it includes the ambient luminance. However, entering the ambient luminance value as 0 cd/m2 can effectively exclude the ambient luminance from the luminance measurement. Note that none of inequalities used in RadiCS includes an equality sign although every judgement condition in JESRA includes it.
RadiCS 97 Index INDEX A AAPM 77, 83 Acceptance Test Carrying Acceptance Test 34 How to Customize the Acceptance Testing Setup 30 How to Display a Pattern for User Setup 32 How to Edit a Check Point 32 ACR 77, 85 B C I J L M CIE 79 Constancy Test E G H Diagnostics 45 DICOM 79 DIN 78, 81 DMG QC Manual 79, 95 O P History Delete 60 Detail Display 59 Export 60 History List 56 History Search 57 Monitor Identification 60 Print 57, 59 Protect/Unprotect 59 Reference 57 Report Generator 59 Research Condit
TG-18 Citation: Samei E, Badano A, Chakraborty D, Compton K, Cornelius C, Corrigan K, Flynn MJ, Hemminger B, Hangiandreou N, Johnson J, Moxley M, Pavlicek W, Roehrig H, Rutz L, Shepard J, Uzenoff R, Wang J, and Willis C. Assessment of Display Performance for Medical Imaging Systems, Report of the American Association of Physicists in Medicine (AAPM) Task Group 18, Medical Physics Publishing, Madison, WI, AAPM On-Line Report No. 03, April 2005.