User`s manual
3. Advanced Settings (Validation, Manual Adjusment)
41
Color Slide bar to the left Slide bar to the right
R to magenta to yellow
G to yellow to cyan
B to cyan to magenta
C to green to blue
M to blue to red
Y to red to green
Tips
• The manual adjustment palette can move and be adjusted with the image.
• The values displayed on the left of the functions, such as white point, brightness, black level, 6
colors adjustment are reference values.
Note
• The monitor gamma adjustment varies depending on the monitor.
• The profile does not reflect the result of the 6 colors adjustment.
• If 6 colors adjustment is performed on CG18, CG19 or CG21, select "Yes" in the 6 color
adjustment window before performing.
2
Measure the monitor
After adjusting manually, click [OK] of the manual adjustment palette.
The manual adjustment palette will close and the measurement window will appear.
Follow the window instructions to measure the monitor.
3
Confirm the result and generate a profile
When the measurement has completed, the adjustment result window will appear. Confirm the
measurement result and click [Save] to generate a profile.
Profile can be edited.
Click [OK].
Note
• If the black level became to minus value and the adjustment failed, an error message is displayed.
Follow the indication displayed in the window and adjust the monitor from the beginning again.
It's because the measurement device could not have been attached to the monitor closely or light
could have seeped through to the sensor of the measurement device during initialization.
4
Save a profile
The above window will appear. Click [Quit].