User`s manual

38
Chapter 4 SelfCorrection
Setting SelfCorrection Schedule
After executing SelfCorrection or monitor adjustment using software (ColorNavigator or ColorNavigator
Elements), set when to execute the next SelfCorrection.
Setting Range
50 hours to 500 hours (Recommended: 200 hours)
Procedure
1. Choose “SelfCorrection” from the Adjustment menu, and press
.
2. Choose “Schedule” from “SelfCorrection”, and press
.
3. Choose “Interval” from “Schedule”, and press
.
4. Select Interval with
or , and press .
5. Press
to exit.
Note
SelfCorrection can also be executed when there are no computer signals being input.
4-2. Executing SelfCorrection
SelfCorrection can be executed manually regardless of the schedule.
Procedure
1. Choose “SelfCorrection” from the Adjustment menu, and press
.
2. Choose “Execute” from “SelfCorrection”, and press
.
The built-in correction sensor is activated, and SelfCorrection is executed.
Attention
After selecting “Execute”, aging may be executed before the built-in correction sensor is executed.