Specifications

CALIBRATION 51
12 Follow the instructions on the ColorCal Scanning Instructions page to scan the ColorCal
Calibration page.
Wait for the scan job to be completed.
13 Press Measure.
The EX700i retrieves the scan data and calculates the measurements.
14 When the measurement is complete, press Test Print to print a Test Page.
You can examine the Test Page to decide if the calibration is successful.
15 If the calibration is successful, press Apply to save the measurements.