User`s guide

Testing in Timed Mode
Follow steps in “Testing in Steady Mode” to purge your dispensing tip with fluid.
Change to TIMED mode.
Set time to 000.0250 seconds.
Rest the dispense tip on the Dot Standards sheet.
Press the foot pedal (or finger switch) to activate the dispense cycle. Note: the pedal only needs
to be pressed for a moment. The complete time will run once you activate the dispense cycle.
Changing the time changes the deposit size. If your dot is too small, increase the dispense
time to increase the deposit size. If your dot is too large, decrease the dispense time to
decrease your deposit size.
Testing in Teach Mode
Follow steps in “Testing in Steady Mode” to purge your dispensing tip with fluid.
Change to TEACH mode by going to the menu and scrolling to TEACH.
Rest the dispense tip on the Dot Standards sheet.
Press the foot pedal to establish your dot size. Depressing the foot pedal repeatedly will have a
cumulative effect on the dot size and dispense time (Note: Data time control knob will not work).
Continue to press the foot pedal until you achieve the correct deposit size.
If needed, press CLEAR to start over.
Save the dispense time setting you just created by pressing SAVE. This automatically exits
TEACH mode and returns you to TIMED mode.
If you have any questions at this point, please call us now.
In the USA, call 800-556-3484 between 8:00 a.m. and 5:30 p.m. Eastern time.
In Europe, call +44 (0) 1582 666334.
In India, call +91 80 4021 3600.
In Asia, call +86 (21) 3866 9006.
In all other areas, call your authorized EFD distributor or +1-401-431-7000.
Initial Test
Procedures
17
www.nordsonefd.com info@nordsonefd.com USA & Canada 800-556-3484 Europe +44 (0) 1582 666334 India+ 91 80 4021 3600 Asia +China: +86 (21) 3866 9006
Sales and service of Nordson EFD dispensing systems are available worldwide.