User`s guide
Auxiliary Port
The auxiliary port, a 6 pin mini DIN connection, is located on the top row of ports. It is designed to
power the Ultimus accessories such as the workstation lamp.
Note: Turn off Ultimus unit before connecting workstation lamp accessory (#7017122).
The auxiliary port is interchangeable with the foot pedal / finger switch port.
Foot Pedal / Finger Switch Port
The switch port, a 6 pin mini DIN connection, is located on the top right corner of the Ultimus. It is
used to connect the foot pedal or finger switch that activates the dispense cycle. The switch port is
interchangeable with the auxiliary port, described above. Be sure to align the foot pedal or finger
switch cord connector so that arrow on the connector is on top.
Exhaust Port
The exhaust port is the larger 8 mm pneumatic fitting located at the bottom left corner of the
Ultimus workstation.
• If needed, you can connect a cleanroom filter muffler to filter output air to meet Fed 209-B
(0.5 micron particulates). Specify #7017049.
• Caution: Use EFD filter/muffler #7016875 or wear adequate ear protection when operating the
vacuum in close proximity for a prolonged period of time.
Air Output Port
The Ultimus features two air output ports to provide regulated air pressure to the syringe barrel. Both
ports have quick-connect fittings and are interchangeable, sharing the same air source.
• You can connect to the port on the front panel or the back panel, whichever is most convenient.
• On the back panel, the air output is the center port on the bottom row.
• Push in the black quick-connect from the EFD adapter assembly to the air output port and twist to
lock. Note: air does not flow through the port until a quick-connect is installed.
Air Input Port
The air input port is the smaller pneumatic fitting (6 mm) located at the bottom right corner of the
Ultimus workstation.
• Push one end of the air input hose into the input fitting on the back of the Ultimus. An 8 ft (2.4 mm)
air hose is provided.
• Connect the other end of the air input hose to your plant air supply. A standard 1/4-inch NPT
fitting is included with your unit.
Note: clean, dry filtered factory air is required to meet warranty. If your air supply is not filtered,
order the five-micron filter regulator specify #7002002.
Warning! Bottled nitrogen can be used. If high pressure bottled air or nitrogen is used, a high
pressure regulator must be installed on the bottle and set at 100 psi (6.9 bar) maximum. In this
instance, the #7002002 filter regulator is not required.
Caution: A pressure regulator of 0 to 100 PSI is intended to be provided before connecting unit to
the air supply.
Features &
Controls:
Back Panel
12