User Guide
Barrel Select
In both PROGRAMMED DISPENSE and MANUAL mode, pressing the top Single Barrel Select
button activates the screen for the barrel size menu and pre-set dispense rates.
• Press the F1 button to set barrel size. Use the (+) or (-) buttons to scroll through the various
size options.
• Press the F3 button to set Dispense Rate. Select Low, Medium, or High.
The dual barrel select button is not active.
Reset
In PROGRAMMED DISPENSE mode, press the Reset button to reset a range of variables,
depending on the screen selected. Reset choices include syringe barrel volume level to 100%,
Dispense Volume to 0cc, Pause to 0 ms, Pullback to 0 steps, and Dispense Rate to the lowest
minimum value possible, depending on the selected syringe barrel size. (Note: to reset Shot Count,
use the F4 button.)
In MANUAL mode, the Reset button is not active.
Home (F1)
In both PROGRAMMED DISPENSE and MANUAL mode, pressing the F1 button activates the
“Retract to Home” screen. This step is required to remove the syringe barrel and/or when you want
to prepare the unit for storage or shipping.
• Choose Yes by pressing the F3 button. This will fully retract the plunger.
• Choose No by pressing the F4 button. This will return you to the Home screen.
Jog (F2)
In PROGRAMMED DISPENSE mode, pressing the F2 button activates the Jog screen. Jog allows
you to make small adjustments to the position of the plunger as it extends forward. Press and hold
F2 until the “g” of Jog is highlighted. Use the FORWARD and BACK buttons to move the plunger
slightly in either direction as needed.
In MANUAL mode, the F2 button is not active.
Features &
Controls
10