N2891A Performance Verification Procedures
N2891A Performance Verification Procedures
The following procedure can be used to test and verify the DC differential gain
accuracy and bandwidth of the N2891A high voltage differential probe. Please
note that these procedures do not indicate that these characteristics are
warranted. Instead, these procedures are meant to give you an idea of how your
probe performs.
Required Test Equipment
DC Differential Gain Accuracy
WARNING: Generator produces hazardous voltages. To avoid shock, do
not touch exposed metal parts after generator output is enabled.
WARNING: These procedures require the application of high voltage to
the inputs of the N2891A probe. Because this adapter has exposed metal
surfaces, only qualified personnel should perform any testing with
voltage levels exceeding 30 Vrms. All pertinent safety rules and
guidelines for elevated voltage measurements should be followed and
adhered to.
1 Set the volts/division on channel 1 of the oscilloscope to 20 mV/div. Set the
seconds/div to 200 s and the acquisition mode to average 32.
Description Critical Specifications Recommended Model
Part Number
Functions
Digitizing Oscilloscope Bandwidth: >50 MHz
1 M/50 selectable input
Agilent MSO9254A Display probe output
Signal Generator
Precision DC voltage source
Amplitude accuracy: less than or
equal to 0.25%
1 M/50 selectable load
Sine wave greater than or equal
to 25 MHz
Fluke 9500B Signal source for DC
gain and bandwidth
BNC Adapter BNC (f) to Dual Banana (m)
Adapter
Agilent 1251-2277 Interconnection
between probe and
generator
50 BNC Feed Through
Adapter
50 precision feed through Agilent 11048C Termination between
probe and calibrator
for bandwidth
verification