Product Manual
SE-601 Series Electrocardiograph User Manual System Setup
- 88 -
To Set Remote IP
Set the REMOTE IP item to the local IP of Smart ECG Viewer.
To Set Local IP
Set the first three sections of the Local IP item to the first three sections of the local IP of Smart
ECG Viewer. The last section of the LOCAL IP item can be set at random, but it can’t be the
same as the last section of the local IP of Smart ECG Viewer.
To Set Gateway
Set the first three sections of the Gateway item to the first three sections of the local IP of Smart
ECG Viewer. The last section of the Gateway item must be set to 001.
To Set Subnet Mask
You can set Subnet Mask in the Transmission Setup window.. Subnet Mask must be set to
255.255.255.000.
After setup, press Enter to confirm. Then the System Setup screen appears.
For SE-601A, if you press Esc to exit the Transmission Setup window after setup, a hint will
pop up to prompt you to save these modifications.
9.6 Display & Sound Setup
Press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor to Display&Sound on the System
Setup screen, and then press Enter to open the Display&Sound Setup window.
In the Display&Sound Setup window, you can set Brightness, Display Colors, Antialising, Key
Volume, Hint Volume, QRS Volume and Notify Volume.
Figure 9-13 SE-601B Display & Sound Setup Window