SE-601 Series Electrocardiograph User Manual System Setup
- 86 -
Figure 9-10 SE-601A Lead Setup Window
After setup, press Enter to confirm. Then the System Setup screen appears.
For SE-601A, if you press Esc to exit the Lead Setup window after setup, a hint will pop up to
prompt you to save these modifications.
9.4.1 Setting Rhythm Lead1/2/3
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 item in
the Lead Setup window, and then press F1 or F2 to select an option.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3
item in the Lead Setup window, and then press F3 or F4 to select an option.
The rhythm lead can be one of 12 standard leads: І, П, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5,
or V6.
In the AUTO mode, when Auto Style is set to 3×4+1r or 6×2+1r, the rhythm lead selected in the
Rhythm Lead1 item will be printed in the ECG reports; when Auto Style is set to 3×4+3r, 3
rhythm leads selected respectively in the Rhythm Lead1/2/3 item will be printed in the ECG
reports.
In the RHYT mode, when Rhythm Style is set to Single Lead, 60s waveform of the rhythm lead
selected in the Rhythm Lead1 item will be printed in the ECG reports; when Rhythm Style is
set to Three Lead, 20s waveform of three rhythm leads selected respectively in the Rhythm
Lead1/2/3 item will be printed in the ECG reports.
In the R-R mode, the R-R analysis report of the rhythm lead selected in the Rhythm Lead1 item
will be printed.
9.4.2 Setting Lead Sequence
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Lead Sequence item in the
Lead Setup window, and then press F1 or F2 to select Standard or Cabrera.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Lead Sequence item in
the Lead Setup window, and then press F3 or F4 to select Standard or Cabrera.
Lead Sequence Lead group 1 Lead group 2 Lead group 3 Lead group 4
Standard
І, II, ІІІ aVR, aVL, aVF V1, V2, V3 V4, V5, V6
Cabrera
aVL, І, -aVR II, aVF, ІІІ V1, V2, V3 V4, V5, V6