When Auto Style is set to 6×2+1r, ECG waves of 12 leads are printed in 2 groups of 6, and the
ECG wave of the rhythm lead selected in the Rhythm Lead1 item is on the bottom of the ECG
report.
When Auto Style is set to 6×2 adjust, ECG waves of 12 leads are printed in 2 groups of 6. 6×2
adjust means that the time period can be adjusted according to Sample Time, which is different
from other styles. For example, when Sample Time in the Work Mode Setup window is set to
12s, each lead group is printed for about 6 seconds in this style.
NOTE: The ECG reports of the 6×2 adjust style can only be printed, and can not be
saved or transmitted. 6×2 adjust only supports the real-time sampling mode.
9.1.5 Specifying Sampling Mode
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Sample Mode item in the
Work Mode Setup window, and then press F1 or F2 to select an option.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Sample Mode item in
the Work Mode Setup window, and then press F3 or F4 to select an option.
When Sample Mode is set to Pre-Sample, 10s ECG data sampled before pressing the
START/STOP key will be printed out.
When Sample Mode is set to Real-time Sample, 10s ECG data sampled from the time of
pressing the START/STOP key will be printed out.
When Sample Mode is set to Period Sample, first you should set Period Interval and Period
Duration in the Date & Time Setup window. For example, if Period Interval is set to 2 minutes,
and Period Duration is set to 24 minutes, after pressing the START/STOP key, the printing
will be performed every two minutes and come to 12 times.
NOTE:
1. There is no saving, copying or transmitting in the Period Sample mode.
2. ID and patient information will not be changed while carrying out the period printing.
3. Period Duration must be an integral multiple of Period Interval, or else the settings
will not be effective.
When Sample Mode is set to Trigger Sample, after pressing the START/STOP key, if
Arrhythmia ECG data, including Asystole, Ventricular Fibrillation/Ventricular Tachycardia,
5>PVCS>=3, Paired PVCS, Bigeminy, Trigeminy, R ON T, single PVC and Missed Beat, is
detected during the learning course, printing will be triggered automatically.