Product Manual
SE-601 Series Electrocardiograph User Manual System Setup
- 75 -
9.1.2 Specifying Manual Style
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Manual Style item in the
Work Mode Setup window, and then press F1 or F2 to select 3 channel or 6 channel.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Manual Style item in
the Work Mode Setup window, and then press F3 or F4 to select 3 channel or 6 channel.
When Manual Style is set to 3 channel, ECG waves of 3 leads are displayed and printed
simultaneously in the MANU mode.
When Manual Style is set to 6 channel, ECG waves of 6 leads are displayed and printed
simultaneously in the MANU mode.
9.1.3 Specifying Rhythm Style
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Rhythm Style item in the
Work Mode Setup window, and then press F1 or F2 to select Single Lead or Three Lead.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Rhythm Style item in
the Work Mode Setup window, and then press F3 or F4 to select Single Lead or Three Lead.
When Rhythm Style is set to Single Lead, in the RHYT mode, 60s ECG waves of the appointed
single rhythm lead will be printed.
When Rhythm Style is set to Three Lead, in the RHYT mode, 20s ECG waves of three
appointed rhythm leads will be printed.
9.1.4 Specifying Auto Style
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Auto Style item in the Work
Mode Setup window, and then press F1 or F2 to select an option.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Auto Style item in the
Work Mode Setup window, and then press F3 or F4 to select an option.
When Auto Style is set to Off, no ECG wave is printed.
When Auto Style is set to 3×4, ECG waves of 12 leads are printed in 4 groups of 3.
When Auto Style is set to 3×4+1r, ECG waves of 12 leads are printed in 4 groups of 3, and the
ECG wave of the rhythm lead selected in the Rhythm Lead1 item is on the bottom of the ECG
report.
When Auto Style is set to 3×4+3r, ECG waves of 12 leads are printed in 4 groups of 3, and ECG
waves of the three rhythm leads selected respectively in the Rhythm Lead1/2/3 item are on the
bottom of the ECG report.
When Auto Style is set to 6×2, ECG waves of 12 leads are printed in 2 groups of 6.