SE-601 Series Electrocardiograph User Manual Managing Files
- 65 -
NOTE: If you don’t set the transmission mode before transmission, the following hint
will pop up.
1) When the main screen1 is displayed, press the function key F5 below to
open the main screen2.
2) Press the function key F1 below Setup to open the System Setup screen.
3) Press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor. When the cursor is on
Transmission, press Enter to open the Transmission Setup window.
4) For SE-601B/C, press F1 or F2 to select Ethernet. For SE-601A, press F3 or F4 to
select Ethernet.
5) Press Tab or Shift + Tab to move the cursor to the Remote IP item. Set the
REMOTE IP item to the local IP of the PC:.
For example, to view the local IP of the PC: Select Start -> Run, and input "cmd"
in the text box. Click on OK, and input "ipconfig" and press Enter, Then, the IP
address of the PC shows as follows:
6) Press Tab or Shift + Tab to move the cursor to the Local IP item. Set the first three
sections of the Local IP item to the first three sections of the local IP of the PC:. The
last section of the LOCAL IP item can be set at random, but it can’t be the same as
the last section of the local IP of the PC:.
7) Press Tab or Shift + Tab to move the cursor to the Gateway item. Set the first three
sections of the Gateway item to the first three sections of the local IP of the PC. The
last section of the Gateway item must be set to 001. Subnet Mask must be set to
255.255.255.000.