Product Manual

SE-601 Series Electrocardiograph User Manual Managing Files
- 64 -
1) When the main screen1 is displayed, press the function key F5 below to
open the main screen2.
2) Press the function key F1 below Setup to open the System Setup screen.
3) Press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor. When the cursor is on
Transmission, press Enter to open the Transmission Setup window.
4) For SE-601B/C, press F1 or F2 to select UART. For SE-601A, press F3 or F4 to
select UART.
5) Press Enter to confirm.
6) Press Esc or the function key F5 below to return to the main screen1.
5. Press the function key F5 below to open the main screen2. Press the function key
F2 below File to display the File Manage screen1.
6. Press the function key F3 below Trans All to transmit all the files to the PC.
7. Or press F1, F2, Shift + F1 or Shift + F2 to highlight a file, and then press the function key
F4 below Select to select the file and open the File Manage screen3. Press the function key
F4 below Trans to transmit the selected file to the PC.
8.2.2 Transmitting Files through the Net Port
1. Connect the PC to the electrocardiograph with an Ethernet cable recommended by the
manufacturer. If the wireless AP transmission is used, connect the PC to one wireless AP, and
connect the electrocardiograph to the other wireless AP. Only the wireless AP recommended
by the manufacturer can be used.
2. Log into the Smart ECG Viewer software.
NOTE: If you don’t connect the net port of the electrocardiograph to the PC (or if you
don’t connect the electrocardiograph/PC to the wireless AP), or you do not log
into Smart ECG Viewer before transmission, the following hint will pop up.
3. Set the transmission mode and the IP address.