Product Manual

SE-601 Series Electrocardiograph User Manual Printing ECG Reports
- 47 -
After setup, press Esc or the function key F5 below to exit the System Setup
screen.
2. When the main screen1 is displayed, press the MODE key to select the MANU mode.
Press F2 to switch the gain. Press F3 to select a paper speed. Press F4 to set the EMG filter
and the Lowpass filter.
3. Press Tab to select the lead group to be printed.
4. Press the START/STOP key to print an ECG report.
5. Press the START/STOP key to stop printing the report.
7.3 Rhythm Mode
Operation Method:
1. To set Rhythm Style, Rhythm Lead1/2/3 or Lead Sequence (Optional)
When the main screen1 is displayed, press the function key F5 below to open the
main screen2. Press the function key F1 below Setup to display the System Setup screen
and the cursor is on Work Mode. Press Enter to open the Work Mode Setup window.
1) For SE-601B/C, press F1 or F2 to select RHYT. For SE-601A, press F3 or F4 to
select RHYT.
2) For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Rhythm Style
item, and then press F1 or F2 to select a style.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Rhythm
Style item, and then press F3 or F4 to select a style.
3) Press Enter to confirm.
SE-601B / SE-601C Work Mode Setup Window
SE-601A Work Mode Setup Window
4) On the System Setup screen, press F1, F2, F3, F4, Tab or Shift + Tab to move the
cursor. When the cursor is on Lead, press Enter to open the Lead Setup window.