Product Manual
SE-601 Series Electrocardiograph User Manual Operation Preparations
- 37 -
4.5.2 Disposable Electrodes
Disposable Electrode Alligator Clip
Disposable electrodes must be used together with alligator clips.
The electrodes’ positions on the body surface are shown in the following table and figures.
American
Label
European
Label
Electrode Placement
RA R Right deltoid
LA L Left deltoid
RL N or RF
Above the right ankle (Alternate placement, upper leg as close to
torso as possible)
LL F
Above the left ankle (Alternate placement, upper leg as close to torso
as possible)
V1 C1 Fourth intercostal space at the right border of the sternum
V2 C2 Fourth intercostal space at the left border of the sternum
V3 C3 Fifth rib between V2 and V4
V4 C4 Fifth intercostal space on the left midclavicular line
V5 C5 Left anterior axillary line at the horizontal level of V4
V6 C6 Left midaxillary line at the horizontal level of V4