Product Manual

SE-601 Series Electrocardiograph User Manual Technical Specifications
- 118 -
Multichannel crosstalk 0.5mm
AC Filter: On/Off
DFT Filter: 0.05Hz / 0.15Hz / 0.25Hz / 0.32Hz / 0.5Hz /
0.67Hz
EMG Filter: 25Hz/35Hz/45Hz/OFF
Filter
LOWPASS Filter:150Hz/100Hz/75Hz
CMRR
115dB
Sampling Frequency
1000 Hz
Pacemaker Detection
Amplitude ±2 to ±700 mV
Width 0.1 to 2.0 ms
Sampling Frequency 10,000 /sec/channel
External Input/Output (Optional)
Input
100k; Sensitivity 10mm/V±5%;
Single ended
Output
100; Sensitivity 1V/mV±5%;
Single ended
NOTE:
1. Test the accuracy of input signal reproduction according to the methods described in
clause 4.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001, and the result complies with the
clause 3.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001.
2. Overall System Error is tested using the method described in AAMI EC11 3.2.7.1.
Overall System Error is not more than ±5%.
3. Frequency Response is tested using the method described in AAMI EC11 3.2.7.2
methods A and D.