Product Manual
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically
- 86 -
In the Transmission Setup window of the electrocardiograph, set Transmission Mode to
Ethernet. Set the Server IP item to the IP of the PC. Set the first three sections of the Local
IP item to the first three sections of the IP of the PC. The last section of the Local IP item is
not the same as the last section of the IP of the PC. Set the first three sections of the Gateway
item to the first three sections of the IP of the electrocardiograph. The last section of the
Gateway item must be set to 1. Commonly, Subnet Mask is set to 255.255.255.0.
8) Press , or press Tab or Shift + Tab to move the cursor to the OK button, and
then press .
9) Press ESC to return to the main screen.
3. Log into the Smart ECG Viewer software.
4. In the auto or rhythm mode, ECG data will be transmitted through the net cable automatically
after an ECG report is printed out. In the off mode, ECG data will be transmitted through the
net cable automatically after the START/STOP key is pressed.