Product Manual
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically
- 85 -
6) Press Tab or Shift + Tab to move the cursor to the Local IP item.
For the cross-network transmission, set the first two sections of the Local IP item to the first
two sections of the IP of the PC. Set the third section of the Local IP item to the network
segment of the electrocardiograph which depends on the configuration of Router. The last
section of the Local IP item can be set at random.
For the same network transmission, set the first three sections of the Local IP item to the first
three sections of the IP of the PC. The last section of the Local IP item can be set at random,
but it can’t be the same as the last section of the IP of the PC.
7) Press Tab or Shift + Tab to move the cursor to the Gateway item. Set the first three
sections of the Gateway item to the first three sections of the IP of the electrocardiograph.
The last section of the Gateway item must be set to 1.
NOTE: For more information on configuring network settings, see your Network
Administrator.
Take the same network transmission for example:
View the IP of the PC.