Product Manual

SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically
- 84 -
5. In the auto or rhythm mode, ECG data will be transmitted through the serial cable
automatically after an ECG report is printed out. In the off mode, ECG data will be
transmitted through the serial cable automatically after the START/STOP key is pressed.
9.1.2 Transmitting ECG Data in DAT Format through Net Cable
1. Connect the electrocardiograph to the network of the PC with an Ethernet cable recommended
by the manufacturer.
If the wireless AP transmission is used, connect the PC to the server wireless AP, and connect
the electrocardiograph to the client wireless AP. Only the wireless AP recommended by the
manufacturer can be used.
NOTE:
1. For details on configuring the wireless AP, please refer to the user manual
delivered with the wireless AP.
2. There should be no shield in or around the room where the wireless AP is used,
or else the wireless transmission may fail.
2. Set Auto Transmission to On and Transmission Mode to Ethernet in the Transmission
Setup window. Set the IP addresses.
1) When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen.
2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to
Transmission, press to open the Transmission Setup window.
3) Press Tab or Shift + Tab to move the cursor to the Auto Transmission list box, and then
press the Up or Down arrow to select On.
4) Press Tab or Shift + Tab to move the cursor to the Transmission Mode list box, and then
press the Up or Down arrow to select Ethernet.
5) Press Tab or Shift + Tab to move the cursor to the Server IP item. Set the Server IP
item to the IP of the PC.