SE-1200 Series Electrocardiograph User Manual Printing ECG Reports
- 66 -
After setup, press ESC to return to the main screen.
2. When the main screen is displayed, press the MODE key to select the R-R analysis mode.
Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter.
3. Press the START/STOP key to begin to count the sampling time. When the sampling time
reaches 180s, it begins to print an ECG report.
4. It will stop automatically after a complete R-R analysis report is printed, or press the
START/STOP key to stop printing the ECG report.
NOTE: In the R-R analysis mode, you can not set the speed. The constant speed is
25mm/s and the printing speed is 5mm/s, because in the R-R analysis mode, the
ECG wave length is compressed to one fifth of the original wave length.
8.5 Copy Printing
In the auto mode, pressing the COPY key can print the ECG report which was printed out last
time. Pressing the START/STOP key can stop printing the ECG report.