Product Manual
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports
- 64 -
After setup, press ESC to return to the main screen.
2. When the main screen is displayed, press the MODE key to select the rhythm mode. Press
F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the
Lowpass filter.
3. Press the START/STOP key to begin to count the sampling time. When the sampling time
reaches 60s in the Single Lead style or 20s in the Three Leads style, it begins to print an
ECG report.
4. It will stop automatically after printing a complete report of rhythm-lead ECG waveforms. Or
press the START/STOP key again to stop printing the ECG report.
8.4 R-R Analysis Mode
Operation Method:
1. To set Mode Options or Rhythm Lead1
When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen. The cursor is on Work Mode. Press to open the Work
Mode Setup window.
1) Press Tab or Shift + Tab to move the cursor to R-R, press Spacebar to select it, and then
a check mark √ appears in the box before R-R.