Product Manual

SE-1200 Series Electrocardiograph User Manual Entering Patient Information
- 47 -
Chapter 6 Entering Patient Information
After the electrocardiograph is turned on, the main screen appears.
6.1 Entering Patient ID
1. By default, the system generates the patient ID automatically. The range of the patient ID is 0
~ 1999, 999, 999.
2. Or the system can generate the patient ID according to the time when you press the
START/STOP key to print an ECG report.
The procedures are as follows:
1) Press the function key F6 below Setup to open the System Setup screen.
2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When
the cursor is on Patient Info, press .
3) Press Tab or Shift + Tab to move the cursor to ID Mode, and then press the Up or Down
arrow to select Time.
4) Press , or press Tab or Shift + Tab to move the cursor to the OK button, and then
press
.
5) Press ESC to return to the main screen.
Then the system will generate the patient ID according to the time when you press the
START/STOP key to print an ECG report.
3. Or you can enter the patient ID manually.
The procedures are as follows:
1) Press the function key F6 below Setup to open the System Setup screen.
2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When
the cursor is on Patient Info, press
.