Product Manual

SE-1200 Series Electrocardiograph User Manual Operation Preparations
- 43 -
NOTE: Long-time measurement with a strong negative pressure on the suction bulb may
cause reddening of the skin. When using the electrode on small children or
patients with delicate skin, squeeze the suction ball lightly.
Limb Electrode Connection:
1) Ensure that the electrodes are clean;
2) Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to
the corresponding electrodes according to the colors and identifiers;
3) Clean the electrode area which is a short distance above the ankle or the wrist with 75%
alcohol;
4) Daub the electrode area on the limb with gel evenly;
5) Place a small amount of gel on the metal part of the limb electrode clamp;
6) Connect the electrode to the limb, and make sure that the metal part is placed on the
electrode area above the ankle or the wrist;
7) Attach all limb electrodes in the same way.
4.5.2.2 Disposable Electrodes
CAUTION
The disposable electrodes can only be used for one time.
Disposable Electrode Alligator Clip
Disposable electrodes must be used together with alligator clips.