SE-1200 Series Electrocardiograph User Manual Operation Preparations
- 42 -
4.5.2 Electrode Placement
Two kinds of electrode can be used, one is the reusable electrode, and the other is the disposable
electrode.
4.5.2.1 Reusable Electrodes
Reusable electrodes include limb electrodes and chest electrodes, as the following figures show.
Chest Electrode
Limb Electrode
Chest Electrode Connection:
1) Ensure that the electrodes are clean;
2) Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to
the corresponding electrodes according to the colors and identifiers;
3) Clean the electrode area on the chest surface with 75% alcohol;
4) Daub the round area of 25mm in diameter on each electrode site with gel evenly;
5) Place a small amount of gel on the brim of the chest electrode’s metal cup;
6) Place the electrode on the chest electrode site and squeeze the suction bulb. Unclench it
and the electrode is adsorbed on the chest;
7) Attach all chest electrodes in the same way.