SE-1200 Series Electrocardiograph User Manual Operation Preparations
- 33 -
♦ Battery
When the battery is used, turn on the unit by pressing on the keyboard. Then the
battery indicator ( ) will be lit and the battery symbol will be displayed on the LCD
screen. Because of the consumption during the storage and transport course, the battery
capacity may not be full. If the symbol and the hint information Battery Weak are
displayed, which means the battery capacity is low, please recharge the battery first.
CAUTION
1. If the electrocardiograph is turned off automatically because of low battery capacity,
the settings may not be saved.
2. If the battery has been fully charged and requires recharging after printing only a few
ECGs, consider replacement.
3. The use of electrocardiograph accessories (such as barcode reader) will deplete
battery power at a faster rate. The battery will require more frequent charging if these
accessories are used with the electrocardiograph.
When the battery is fully charged, SE-1200 can work normally about 4 hours, and print
about 300 ECG reports of 3×4+1R in the auto mode; SE-1200 Express can work normally
about 5 hours, and print about 350 ECG reports of 3×4+1R in the auto mode.
Please refer to the maintenance section for how to recharge the battery. When the battery is
being recharged, the electrocardiograph can be powered by the mains supply.
WARNING
Potential equalization conductor of the unit should be connected to the potential
equalization bus bar of the electrical installation when necessary.