Product Manual

SE-1200 Series Electrocardiograph User Manual Introduction
- 20 -
D Heat Emission Hole Path for internal heat emission
E
Label
Position for product information label
1) Battery Compartment
The rated voltage and the rated capacity of the battery pack are as follows:
SE-1200 Express: Rated Voltage: 14.8V; Rated Capacity: 4400mAh
SE-1200: Rated Voltage: 14.8V; Rated Capacity: 2200mAh
WARNING
1. Improper operation may cause the battery to be hot, ignited or exploded, and it may
lead to the decrease of the battery capacity. Therefore, it is necessary to read the
user manual carefully and pay more attention to warning messages.
2. When leakage or foul smell is found, stop using the battery immediately. If your skin
or cloth comes into contact with the leakage liquid, cleanse it with clean water at once.
If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with
clean water first and go to see a doctor immediately.
3. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and
specification must be used.
4. Only when the device is off can the battery be installed or removed.
NOTE: If the battery has not been used for two months or more, you should recharge it
before using it again.
2) Fuse
There are two fuses of the same specification installed on the bottom of the main unit. The
specification is: T1AL250VP Ø5×20.
WARNING
Ruptured fuses must only be replaced with those of the same type and rating as the
original.