SE-1200 Series Electrocardiograph User Manual EMC Information
- 162 -
(>95% dip in U
T
)
for 5 sec
(>95% dip in U
T
)
for 5 sec
supply or a battery.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration - electromagnetic immunity - for EQUIPMENT
and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration - electromagnetic immunity
The 12-channel electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or the user of the 12-channel electrocardiograph should assure
that it is used in such an environment.
Immunity
test
IEC/EN 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC/EN
61000-4-6
Radiated RF
IEC/EN
61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3V
rms
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the
12-channel electrocardiograph,
including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Pd 2.1
Pd 2.1 80 MHz to 800 MHz
Pd 3.2 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol: