Product Manual

SE-1200 Series Electrocardiograph User Manual EMC Information
- 160 -
Appendix 2 EMC Information
Guidance and manufacture’s declaration - electromagnetic emissions - for all EQUIPMENT
and SYSTEMS
Guidance and manufacture’s declaration - electromagnetic emission
The 12-channel electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or the user of the 12-channel electrocardiograph should assure
that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11 Group 1
The 12-channel electrocardiograph uses RF
energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class A
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Complies
The 12-channel electrocardiograph is suitable
for use in all establishments, other than
domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.