Product Manual

SE-1200 Series Electrocardiograph User Manual Technical Specifications
- 157 -
Charge mode: Constant current/voltage
SE-1200 Express: Charge current (standard) = 0.14 C
5
A
(600mA)
SE-1200: Charge current (standard) = 0.28 C
5
A (600mA)
Charge voltage (standard) = (16.8-0.1V)
Necessary Charge time:
SE-1200 Express: 9.5 hours
SE-1200: 5 hours
Cycle life 300 times
Fuse:
T1AL250VP Ø5×20
A1.5 Performance Specifications
Recording
Recorder: Thermal dot-matrix recorder
Printing Density
8 dots per mm / 200 dots per inch (amplitude axes)
40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s)
Recorder Paper:
Folded thermal paper: 210mm×295mm×100pages
Folded thermal paper: 215mm×280mm×100pages
(Optional)
Rolled thermal paper: 210mm×30m (Optional)
Effective Width: 203mm
Paper Speed:
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s
±3%
Accuracy of data: ±5% (x-axis), ±5%(y-axis)
HR Recognition
Technique: Peak-peak detection
HR Range: 30 BPM ~300 BPM
Accuracy: ±1 BPM
ECG Unit
Leads: 12 standard leads